Routine Replacement/Adjustment
7
Remove the two countersunk M4 x 10 Phillips-head screws in the Plate Loader module. Refer to
Figure
11.3.
Figure 11.3 Plate Loader Module Securing Screws
1. Securing screws
8
Slide the Plate Loader module out of the Cytometer. Refer to
Figure 11.4 Removing the Plate Loader from the Cytometer

Installation

1
Remove the front cover. Refer to
11-50
b
Front Cover Removal and
Figure
11.4.
Reinstallation.