Instrument Installation
Unpacking the Instrument and Inspecting the Materials for Defects or Omissions [CytoFLEX]
13
Remove the foam support holding the sample loader in the front-right corner of the
instrument.
[CytoFLEX Without Plate Loader Shown]
14
Add Deep Clean solution to the Deep Clean solution bottle. Refer to
Solution
15
Clean the sheath fluid container. Refer to
CHAPTER 10, Cleaning
16
Fill the Sheath fluid container. Refer to
CHAPTER 11, Replacement/Adjustment
WARNING
Risk of chemical injury from bleach. To avoid contact with the bleach, use barrier
protection, including protective eyeware, gloves, and suitable laboratory attire.
Refer to the Safety Data Sheet for details about chemical exposure before using
the chemical.
17
Add 400 mL of 5 to 6% bleach to the waste container.
18
Install the CytExpert software. Refer to
19
Turn on the instrument. Refer to
20
Open the CytExpert software. Refer to
21
Run the System Startup Program. Refer to
Tube Loader
A-10
in
CHAPTER 11, Replacement/Adjustment
Procedures.
Turning On the Instrument
in
CHAPTER 3, Daily
Startup.
Procedures.
Cleaning the 4 L Sheath Fluid Container
Filling the 4 L Sheath Fluid Container [CytoFLEX]
Procedures.
Installing the Software
[CytoFLEX].
in
Logging Into the Software
Running the System Startup Program with the Single
Adding the Deep Clean
in
CHAPTER 3, Daily
Startup.
in
CHAPTER 3, Daily
Startup.
B49006AH
in