Power Source

DANGER
Risk of electric shock and/or instrument damage. Ensure that the power source is
properly grounded. Improper grounding can cause electric shock and damage the
system. Verify that the output voltage of the power outlet conforms to the system
requirements and that a 5 A, time delay, T 5 AL 250 VAC fuse is installed. To
prevent personal injury, Beckman Coulter recommends using a power source
designed to protect against electrical shock.
CAUTION
Possible instrument damage could occur if you use an extension cord or a power
strip to connect the Cytometer. Always plug the Cytometer into a dedicated outlet
with an isolated ground.
The power source requirements are as follows:
• This instrument has been tested to and meets all applicable requirements for CE Marking.
• This instrument complies with the emission and immunity requirements described in IEC
61326-1 and 61326-2-6 series.
• This equipment has been designated and tested to CISPR 11 Class A. In a domestic environment
it may cause radio interference, in which case, you may need to take measures to mitigate the
interference.
• It is advised that the electromagnetic environment should be evaluated prior to operation of
the device.
• Do not use this device in close proximity to sources of strong electromagnetic radiation
(unshielded intentional RF sources), as these may interfere with the proper operation.
• 100-240 volts, 50/60 Hz, 3-wire power cable, well grounded.
• Amperage not less than 10 A.
• The system requires a well-grounded power outlet (150 VA normal, 250 VA max) to provide the
necessary power.
• Distance from system to socket less than 1.5 m.
Power consumption of the Plate Loader is <30 W.
B49006AH