Daily Startup
Pre-Startup Inspection

Check Waste and Reagent Levels [4 L Fluid Containers]

CAUTION
Risk of instrument damage. Do not use a saline-based sheath fluid on the
CytoFLEX instrument. Saline-based sheath fluid could damage instrument
components. Beckman Coulter recommends using CytoFLEX Sheath Fluid or a
similar nonionic sheath fluid to ensure system performance.
1
Examine the sheath fluid and waste containers. Verify that there is sufficient sheath fluid in the
sheath fluid container and that the waste container is empty.
NOTE
When the sheath fluid container is near empty or the waste container is near full, a warning
message is transmitted to the Workstation and audible signals sound as a warning.
b
1. Sheath fluid container
2. Waste container
CAUTION
Risk of instrument damage. Remove the sheath fluid container from the Fluid
Container holder before filling the sheath fluid container to avoid damage to
instrument electronics.
2
If necessary, fill the sheath fluid container with CytoFLEX Sheath Fluid or a similar nonionic
sheath fluid while not exceeding the maximum volume indicated (4 L). Refer to
Sheath Fluid Container [CytoFLEX]
3-2
Sheath
Sheath
Waste
in
CHAPTER 11, Replacement/Adjustment
c
Procedures.
Filling the 4 L
B49006AH