2 Command System
[:SOURce<n>]:FUNCtion:RAMP:SYMMetry
Syntax
[:SOURce<n>]:FUNCtion:RAMP:SYMMetry {<val>|MINimum|MAXimum}
[:SOURce<n>]:FUNCtion:RAMP:SYMMetry? [MINimum|MAXimum]
Description
Set or query the symmetry of the ramp.
Parameter
Name
Type
<val>
Real
<n>
Discrete
Explanation
Symmetry is defined as the percentage that the rising period takes up in the whole period.
t
symmetry=t/T*100%
T
Return Format
The query returns the current symmetry in scientific notation.
MSO2000A/DS2000A Programming Guide
Range
0% to 100%
{1|2}
Default
10%