Measure:setup:max; Measure:setup:mid; Measure:setup:min - Rigol MSO2000A Series Programming Manual

Digital oscilloscope
Hide thumbs Also See for MSO2000A Series:
Table of Contents

Advertisement

2 Command System

:MEASure:SETup:MAX

:MEASure:SETup:MID

:MEASure:SETup:MIN

Syntax
:MEASure:SETup:MAX <max>
:MEASure:SETup:MAX?
:MEASure:SETup:MID <mid>
:MEASure:SETup:MID?
:MEASure:SETup:MIN <min>
:MEASure:SETup:MIN?
Description
Set or query the upper limit, middle value or lower limit of the auto measurement of the analog channels.
The unit is %.
Parameter
Name
Type
<max>
Integer
<mid>
Integer
<min>
Integer
Explanation
The upper limit, middle value and lower limit of the threshold level are expressed in percentage of
the amplitude and are used to define the time, delay and phase parameters. Setting the threshold
will affect all the time, delay and phase parameters. You can also use the
command to select the threshold measurement setting.
The middle value set must be lower than the upper limit currently set and greater than the lower limit
currently set.
Return Format
The query returns an integer.
MSO2000A/DS2000A Programming Guide
Range
7 to 95
6 to 94
5 to 93
Default
90
50
10
:MEASure:SETup:TYPE
RIGOL
2-141