:TRIGger:EDGe:SLOPe
Syntax
:TRIGger:EDGe:SLOPe <slope>
:TRIGger:EDGe:SLOPe?
Description
Set or query the edge type in edge trigger.
Parameter
Name
Type
<slope>
Discrete
Explanation
POSitive: trigger on the rising edge of the input signal when the voltage level meets the preset
trigger level.
NEGative: trigger on the falling edge of the input signal when the voltage level meets the preset
trigger level.
RFALl: trigger on the rising or falling edge of the input signal when the voltage level meets the preset
trigger level.
Return Format
The query returns POS, NEG or RFAL.
:TRIGger:EDGe:SOURce
Syntax
:TRIGger:EDGe:SOURce <src>
:TRIGger:EDGe:SOURce?
Description
Set or query the trigger source of edge trigger.
Parameter
Name
Type
<src>
Discrete
Return Format
The query returns D0, D1, D2, D3, D4, D5, D6, D7, D8, D9, D10, D11, D12, D13, D14, D15, CHAN1,
CHAN2, EXT or ACL.
2-278
Range
{POSitive|NEGative|RFALl}
Range
{D0|D1|D2|D3|D4|D5|D6|D7|D8|D9|D10|D11|D12|
D13|D14|D15|CHANnel1|CHANnel2|EXT|ACLine}