2 Command System
:TRIGger:TIMeout:SLOPe
Syntax
:TRIGger:TIMeout:SLOPe <slope>
:TRIGger:TIMeout:SLOPe?
Description
Set or query the edge type of timeout trigger.
Parameter
Name
Type
<slope>
Discrete
Explanation
: start timing when the rising edge of the input signal passes through the trigger level.
: start timing when the falling edge of the input signal passes through the trigger level.
: start timing when any edge of the input signal passes through the trigger level.
Return Format
The query returns POS, NEG or RFAL.
:TRIGger:TIMeout:TIMe
Syntax
:TRIGger:TIMeout:TIMe <NR3>
:TRIGger:TIMeout:TIMe?
Description
Set or query the timeout time of timeout trigger.
Parameter
Name
Type
<NR3>
Real
Return Format
The query returns the timeout time in scientific notation.
MSO2000A/DS2000A Programming Guide
Range
{POSitive|NEGative|RFALl}
Range
16ns to 4s
Default
POSitive
Default
1us
RIGOL