Display:mpersistence; Display:data - Rigol MSO2000A Series Programming Manual

Digital oscilloscope
Hide thumbs Also See for MSO2000A Series:
Table of Contents

Advertisement

RIGOL

:DISPlay:MPERsistence

Syntax
:DISPlay:MPERsistence <time>
:DISPlay:MPERsistence?
Description
Set or query the menu display time and the unit is s.
Parameter
Name
Type
<time>
Discrete
Return Format
The query returns the menu display time set.

:DISPlay:DATA?

Syntax
:DISPlay:DATA?
Description
Read the bitmap data stream of the image currently displayed.
Explanation
The command is sent from the PC to the instrument through the VISA interface. The instrument responds
to the command and directly returns the bitmap data stream of the image currently displayed to the
buffer area of the PC.
Return Format
Th format of the bitmap data stream:
Component
TMC Blockheader
[1]
Size (length)
N
Example
#9001152054
TMC Blockheader ::= #NX...X is used to describe the
length of the data stream. Wherein, # is the start
denoter of the data stream; N is less than or equal to 9
Explanation
and the N figures following it denotes the length of the
data stream in bytes. For example, #9001152054;
wherein, N is 9 and 001152054 denotes that the data
stream contains 1152054 bytes of effective data.
[1]
Note
: N is the width used to describe the data length in the TMC header. For example, #9001152054.
2-92
Range
{1|2|5|10|20|INFinite}
+2
Default
INFinite
BMP Data
800*480*3+54=1152054
BM...
Specific bitmap data.