Rigol MSO2000A Series Programming Manual page 42

Digital oscilloscope
Hide thumbs Also See for MSO2000A Series:
Table of Contents

Advertisement

RIGOL
:BUS<n>:RS232:ENDian
Syntax
:BUS<n>:RS232:ENDian <endian>
:BUS<n>:RS232:ENDian?
Description
Set or query the endian of data transmission of RS232 decoding on the specified bus.
Parameter
Name
Type
<n>
Discrete
<endian>
Discrete
Return Format
The query returns MSB or LSB.
:BUS<n>:RS232:BAUD
Syntax
:BUS<n>:RS232:BAUD <baud>
:BUS<n>:RS232:BAUD?
Description
Set or query the baud rate of data transmission of RS232 decoding on the specified bus. The default unit
is bps.
Parameter
Name
Type
<n>
Discrete
<baud>
Discrete
Note: When the baud rate is set to USER, you need to set a specific baud rate using the
command.
Return Format
The query returns the baud rate currently set in integer or returns USER.
2-26
Range
{1|2}
{MSB|LSB}
Range
{1|2}
{2400|4800|9600|19200|38400|57600|115200|USER}
2 Command System
Default
--
LSB
Default
--
9600
:BUS<n>:RS232:BUSer