:CHANnel<n>:SCALe
Syntax
:CHANnel<n>:SCALe <scale>
:CHANnel<n>:SCALe?
Description
Set or query the vertical scale of the specified analog channel. The default unit is V.
Parameter
Name
Type
<n>
Discrete
<scale>
Real
Note: The range of the vertical scale is related to the probe ratio and input impedance currently set. For the setting
of the probe ratio, refer to the
the
:CHANnel<n>:IMPedance
Return Format
The query returns the vertical scale in scientific notation.
2-70
Range
{1|2}
When the input impedance is 50Ω
and the probe ratio is 1X:
500μV/div to 1V/div
When the input impedance is 1MΩ
and the probe ratio is 1X:
500μV/div to 10V/div
:CHANnel<n>:PROBe
command.
Default
--
1V
command. For the input impedance, refer to
MSO2000A/DS2000A Programming Guide