[:Trace<N>]:Data:dac16 - Rigol MSO2000A Series Programming Manual

Digital oscilloscope
Hide thumbs Also See for MSO2000A Series:
Table of Contents

Advertisement

RIGOL
[:TRACe<n>]:DATA:DAC16
Syntax
[:TRACe<n>]:DATA:DAC16 volatile,<flag>,<binary_block_data>
Description
Download binary data block into the volatile memory of the built-in signal source.
Parameter
Name
<n>
<flag>
<binary_block_data>
Explanation
This command consists of two parts: the command string ([:TRACe<n>]:DATA:DAC16
volatile,<flag>) and the bianry data (<binary_block_data>).
<flag> denotes the data transmission status and can only be set to END which denotes this data
packet is the last one and the data transmission finishes.
<binary_block_data> denotes the binary data to be downloaded. The range is from 0000 to 3FFF.
The data length is from 2pts (4Bytes) to 16kpts (32kBytes). The binary data block starts with #. For
example, send the command :DATA:DAC16 VOLATILE,END,#532768 binary data
The number "5" behind # denotes that the data length information (32768) occupies 5 characters.
32768 denotes the number of bytes of the binary data. As each waveform point occupies two
bytes, the number of bytes must be an even number.
When the data transmission complte denoter END is recieved, the instrument switches to arbitrary
waveform output automatically.
2-260
Type
Range
Discrete
{1|2}
Discrete
{END}
Refer to the Explanation
2 Command System
Default
1
--
MSO2000A/DS2000A Programming Guide