RIGOL

:MEASure:OVERshoot

Syntax 1
:MEASure:OVERshoot <chan>
Description 1
Enable the overshoot measurement function of the specified channel.
Syntax 2
:MEASure:OVERshoot? <chan>
Description 2
Enable the overshoot measurement function of the specified channel and return the measurement result.
Parameter
Name
Type
<chan>
Discrete
Return Format
The query returns the measurement result in scientific notation.
Example
:MEASure:OVERshoot? CHANnel1
:MEASure:OVERshoot? CHANnel1
You can use the following commands to query the statistic maximum, minimum, current value, average
and standard deviation of the overshoot measurement values of the specified channel waveform. Sending
the following commands will enable the statistic function automatically.
:MEASure:OVERshoot:SMAXimum?
:MEASure:OVERshoot:SMINimum?
:MEASure:OVERshoot:SCURrent?
:MEASure:OVERshoot:SAVerage?
:MEASure:OVERshoot:SDEViation?
2-162
Range
{CHANnel1|CHANnel2|MATH}