Channel<N>:Offset - Rigol MSO2000A Series Programming Manual

Digital oscilloscope
Hide thumbs Also See for MSO2000A Series:
Table of Contents

Advertisement

2 Command System
:CHANnel<n>:OFFSet
Syntax
:CHANnel<n>:OFFSet <offset>
:CHANnel<n>:OFFSet?
Description
Set or query the vertical offset of the specified analog channel. The default unit is V.
Parameter
Name
Type
<n>
Discrete
<offset>
Real
Note: The range of the vertical offset is related to the vertical scale, probe ratio and input impedancecurrently set.
For the vertical scale, refer to the
the
:CHANnel<n>:PROBe
Return Format
The query returns the vertical offset in scientific notation.
MSO2000A/DS2000A Programming Guide
Range
{1|2}
When the input impedance is 50Ω and the
probe ratio is 1X:
500μV/div to 50mV/div: -2V to 2V
51mV/div to 200mV/div: -10V to 10V
205mV/div to 1V/div: -12V to 12V
When the input impedance is 1MΩ and the
probe ratio is 1X:
500μV/div to 50mV/div: -2V to 2V
51mV/div to 200mV/div: -10V to 10V
205mV/div to 2V/div: -50V to 50V
2.05V/div to 10V/div: -100V to 100V
:CHANnel<n>:SCALe
command. For the input impedance, refer to the
Default
--
CHANnel1: 2V
CHANnel2: -2V
command. For the probe ratio, refer to
:CHANnel<n>:IMPedance
RIGOL
command.