RIGOL
[:TRACe<n>]:DATA:LOAD?
Syntax
[:TRACe<n>]:DATA:LOAD? 1
Description
Read the specified data packet in the volatile memory of the built-in signal source.
Parameter
Name
Type
<n>
Discrete
[:TRACe<n>]:DATA:POINts
Syntax
[:TRACe<n>]:DATA:POINts VOLATILE, {<point>|MINimum|MAXimum}
[:TRACe<n>]:DATA:POINts? VOLATILE[,MINimum|MAXimum]
Description
Set or query the number of arbitrary waveform initial points of the specified built-in signal source channel.
Parameter
Name
Type
<point>
Integer
<n>
Discrete
Explanation
This command sets the amplitude of the initial points to 0 automatically.
Return Format
The query returns an integer between 2 and 16384.
Example
:DATA:POINts VOLATILE,200
2-262
Range
{1|2}
Range
2 to 16384
{1|2}
2 Command System
Default
1
Default
2