Rigol MSO2000A Series Programming Manual page 121

Digital oscilloscope
Hide thumbs Also See for MSO2000A Series:
Table of Contents

Advertisement

2 Command System
:FUNCtion:WANalyze:TDISp
Syntax
:FUNCtion:WANalyze:TDISp <bool>
:FUNCtion:WANalyze:TDISp?
Description
Enable or disable the template display; or query the current status of template display.
Parameter
Name
Type
<bool>
Bool
Return Format
The query returns 0 or 1.
:FUNCtion:WANalyze:SETup:SSTart
:FUNCtion:WANalyze:SETup:SSENd
Syntax
:FUNCtion:WANalyze:SETup:SSTart <start>
:FUNCtion:WANalyze:SETup:SSTart?
:FUNCtion:WANalyze:SETup:SSENd <end>
:FUNCtion:WANalyze:SETup:SSENd?
Description
Set or query the screen start point or end point of waveform analysis.
Parameter
Name
Type
<start>
Integer
<end>
Integer
Explanation
The screen start point set should be less than the "screen end-10".
The screen end point set should be greater than the "screen start+10".
Return Format
The query returns an integer.
MSO2000A/DS2000A Programming Guide
Range
{{0|OFF}|{1|ON}}
Range
5 to 685
15 to 695
Default
0|OFF
Default
5
695
RIGOL
2-105