:CHANnel<n>:VERNier
Syntax
:CHANnel<n>:VERNier <bool>
:CHANnel<n>:VERNier?
Description
Enable or disable the fine adjustment function of the vertical scale of the specified analog channel; or
query the current status of the fine adjustment function of the vertical scale of the specified analog
channel.
Parameter
Name
Type
<n>
Discrete
<bool>
Bool
Explanation
By default, the fine adjustment function is disabled. At this point, you can only set the vertical scale in
1-2-5 step; namely 500uV, 1mV, 2mV, 5mV, 10mV...10V. When the fine adjustment function is enabled,
you can further adjust the vertical scale within a relatively smaller range to improve vertical resolution. If
the amplitude of the input waveform is a little bit greater than the full scale under the current scale and
the amplitude would be a little bit lower if the next scale is used, fine adjustment can be used to improve
the amplitude of waveform display to view signal details.
Return Format
The query returns 0 or 1.
2-72
Range
{1|2}
{{0|OFF}|{1|ON}}
Default
--
0|OFF