2 Command System

:ACQuire Commands

:ACQuire:AVERages

:ACQuire:MDEPth
:ACQuire:SRATe?
:ACQuire:TYPE
:ACQuire:AALias
:ACQuire:AVERages
Syntax
:ACQuire:AVERages <count>
:ACQuire:AVERages?
Description
Set or query the number of averages in average acquisition mode.
Parameter
Name
Type
<count>
Discrete
Explanation
You can use the
:ACQuire:TYPE
oscilloscope averages the waveforms from multiple samples to reduce the random noise on the input
signal and improve the vertical resolution. Greater number of averages can lower the noise and increase
the vertical resolution; while at the same time, it will slower the response of the displayed waveform to
the waveform changes.
Return Format
The query returns an integer.
MSO2000A/DS2000A Programming Guide
Range
n
2
(n is an integer from 1 to 13)
command to select the average acquisition mode. In this mode, the
RIGOL
Default
2