Rigol MSO2000A Series Programming Manual page 317

Digital oscilloscope
Hide thumbs Also See for MSO2000A Series:
Table of Contents

Advertisement

2 Command System
:TRIGger:RUNT:WHEN
Syntax
:TRIGger:RUNT:WHEN <when>
:TRIGger:RUNT:WHEN?
Description
Set or query the qualifier of runt trigger.
Parameter
Name
Type
<when>
Discrete
Explanation
NONE: do not set the trigger condition of runt trigger.
GREater (>):trigger when the runt pulse width is greater than the lower limit of pulse width (refer to
the
:TRIGger:RUNT:WLOWer
LESS (<):trigger when the runt pulse width is lower than the upper limit of pulse width (refer to
the
:TRIGger:RUNT:WUPPer
GLESs (<>):trigger when the runt pulse width is greater than the lower limit (refer to
the
:TRIGger:RUNT:WLOWer
the
:TRIGger:RUNT:WUPPer
Note: The lower limit of the pulse width must be lower than the upper limit.
Return Format
The query returns NONE, GRE, LESS or GLES.
:TRIGger:RUNT:WLOWer
Syntax
:TRIGger:RUNT:WLOWer <NR3>
:TRIGger:RUNT:WLOWer?
Description
Set or query the lower limit of the pulse width in runt trigger.
Parameter
Name
Type
<NR3>
Discrete
Return Format
The query returns the lower limit of the pulse width in scientific notation.
MSO2000A/DS2000A Programming Guide
Range
{NONE|GREater|LESS|GLESs}
command).
command).
command) and lower than the upper limit (refer to
command) of pulse width.
Range
When the qualifier is >: 2ns to 4s
When the qualifier is <>: 2ns to 3.99s
RIGOL
Default
NONE
Default
1us