:MEASure:VTOP

Syntax 1
:MEASure:VTOP <chan>
Description 1
Enable the amplitude top value measurement function of the specified channel.
Syntax 2
:MEASure:VTOP? <chan>
Description 2
Enable the amplitude top value measurement function of the specified channel and return the
measurement result.
Parameter
Name
Type
<chan>
Discrete
Explanation
The unit is determined by the unit selected for the source channel (refer to the
command).
Return Format
The query returns the measurement result in scientific notation.
Example
:MEASure:VTOP CHANnel1
:MEASure:VTOP? CHANnel1
You can use the following commands to query the statistic maximum, minimum, current value, average
and standard deviation of the amplitude top measurement values of the specified channel waveform.
Sending the following commands will enable the statistic function automatically.
:MEASure:VTOP:SMAXimum?
:MEASure:VTOP:SMINimum?
:MEASure:VTOP:SCURrent?
:MEASure:VTOP:SAVerage?
:MEASure:VTOP:SDEViation?
2-198
Range
{CHANnel1|CHANnel2|MATH}
2 Command System
Default
--
:CHANnel<n>:UNITs