2 Command System
:TRIGger:IIC:CLEVel
:TRIGger:IIC:DLEVel
Syntax
:TRIGger:IIC:CLEVel <level>
:TRIGger:IIC:CLEVel?
:TRIGger:IIC:DLEVel <level>
:TRIGger:IIC:DLEVel?
Description
Set or query the trigger level of SCL or SDA in I2C trigger. The unit is the same with the current amplitude
unit.
Parameter
Name
Type
<level>
Real
Note: For the VerticalScale, refer to the
the
:CHANnel<n>:OFFSet
Explanation
To set the I2C trigger condition, refer to the
Return Format
The query returns the trigger level in scientific notation.
:TRIGger:IIC:DATA
Syntax
:TRIGger:IIC:DATA <dat>
:TRIGger:IIC:DATA?
Description
Set or query the data value in I2C trigger when the trigger condition is Data or A&D.
Parameter
Name
Type
<dat>
Integer
Explanation
To set the I2C trigger condition, refer to the
Return Format
The query returns an integer.
MSO2000A/DS2000A Programming Guide
Range
When the clock source or data source is CHANnel1 or CHANnel2:
-5×VerticalScale–OFFSet to +5×VerticalScale–OFFSet
When the clock source or data source is any of D0-D15: -20V to 20V
:CHANnel<n>:SCALe
command.
:TRIGger:IIC:WHEN
Range
40
0 to 2
-1
:TRIGger:IIC:WHEN
command. For the OFFSet, refer to
command.