2 Command System
:SAVE:IMAGe:TYPE
Syntax
:SAVE:IMAGe:TYPE <type>
:SAVE:IMAGe:TYPE?
Description
Set or query the picture type of picture storage.
Parameter
Name
Type
<type>
Discrete
Return Format
The query returns BMP24, JPEG, PNG or TIFF.
:SAVE:IMAGe:FACTors
Syntax
:SAVE:IMAGe:FACTors <bool>
:SAVE:IMAGe:FACTors?
Description
Enable or disable the parameter storage function of picture storage; or query whether the parameter
storage function of picture storage is enabled.
Parameter
Name
Type
<bool>
Bool
Explanation
When this function is enabled, a TXT file with the same filename will also be stored when sending
the
:SAVE:IMAGe[:STARt]
hardware and software version numbers as well as the current parameter configurations of the
oscilloscope.
Return Format
The query returns 0 or 1.
MSO2000A/DS2000A Programming Guide
Range
{BMP24bit|JPEG|PNG|TIFF}
Range
{{0|OFF}|{1|ON}}
command to store the picture file. This file includes the model, serial number,
Default
PNG
Default
0|OFF
RIGOL
2-217