2 Command System
:MEASure:VAMP
:MEASure:VAVG
:MEASure:VBASe
:MEASure:VMAX
:MEASure:VMIN
:MEASure:VPP
:MEASure:VRMS
:MEASure:PVRMs
:MEASure:VTOP
:MEASure:MARea
:MEASure:MPARea

:MEASure:SOURce

Syntax
:MEASure:SOURce <sour>
:MEASure:SOURce?
Description
Set or query the signal source of the current measurement parameter.
Parameter
Name
Type
<sour>
Discrete
Explanation
Only channels currently turned on can be selected.
This command sets the signal source of all the measurement parameters except delay and phase.
The delay and phase measurements require two signal sources which can be set by
the
:MEASure:SETup:DSA
and
:MEASure:SETup:PSB
Return Format
The query returns D0, D1, D2, D3, D4, D5, D6, D7, D8, D9, D10, D11, D12, D13, D14, D15, CHAN1,
CHAN2 or MATH.
MSO2000A/DS2000A Programming Guide
Range
{D0|D1|D2|D3|D4|D5|D6|D7|D8|D9|D10|D11|D12|D13|D14|
D15|CHANnel1|CHANnel2|MATH}
and :MEASure:SETup:DSB,
commands respectively.
:MEASure:SETup:PSA
RIGOL
Default
CHANnel1