Page 1
RIGOL User’s Guide MSO4000/DS4000 Series Digital Oscilloscope Mar. 2015 RIGOL Technologies, Inc. 800.544.2843 www.calcert.com sales@calcert.co...
Notices RIGOL products are covered by P.R.C. and foreign patents, issued and pending. RIGOL reserves the right to modify or change parts of or all the specifications and pricing policies at company’s sole decision. Information in this publication replaces all previously corresponding material.
RIGOL Safety Requirement General Safety Summary Please review the following safety precautions carefully before putting the instrument into operation so as to avoid any personal injury or damage to the instrument and any product connected to it. To prevent potential hazards, please use the instrument only specified by this manual.
Page 5
Do not touch exposed junctions and components when the unit is powered. Do Not Operate With Suspected Failures. If you suspect damage occurs to the instrument, have it inspected by RIGOL authorized personnel before further operations. Any maintenance, adjustment or replacement especially to circuits or accessories must be performed by RIGOL authorized personnel.
Page 6
If a battery is supplied, it must not be exposed to high temperature or in contact with fire. Keep it out of the reach of children. Improper change of battery (note: lithium battery) may cause explosion. Use RIGOL specified battery only. Handling Safety.
RIGOL Safety Terms and Symbols Terms Used in this Manual. These terms may appear in this manual: WARNING Warning statements indicate conditions or practices that could result in injury or loss of life. CAUTION Caution statements indicate conditions or practices that could result in damage to this product or other property.
RIGOL Allgemeine Sicherheits Informationen Überprüfen Sie diefolgenden Sicherheitshinweise sorgfältigumPersonenschädenoderSchäden am Gerätundan damit verbundenen weiteren Gerätenzu vermeiden. Zur Vermeidung vonGefahren, nutzen Sie bitte das Gerät nur so, wiein diesem Handbuchangegeben. Um Feuer oder Verletzungen zu vermeiden, verwenden Sie ein ordnungsgemäßes Netzkabel.
Page 9
Wenn Sie am Gerät einen Defekt vermuten, sorgen Sie dafür, bevor Sie das Gerät wieder betreiben, dass eine Untersuchung durch RIGOL autorisiertem Personal durchgeführt wird. Jedwede Wartung, Einstellarbeiten oder Austausch von Teilen am Gerät, sowie am Zubehör dürfen nur von RIGOL autorisiertem Personal durchgeführt werden. Belüftung sicherstellen.
Page 10
Wenneine Batterieverwendet wird, vermeiden Sie hohe Temperaturen bzw. Feuer ausgesetzt werden. Bewahren Sie es außerhalbder Reichweitevon Kindern auf. UnsachgemäßeÄnderung derBatterie (Anmerkung: Lithium-Batterie) kann zu einer Explosion führen. VerwendenSie nur von RIGOL angegebenenAkkus. Sicherer Transport. Transportieren Sie das Gerät sorgfältig (Verpackung!), um Schäden an Bedienelementen, Anschlüssen und anderen Teilen zu vermeiden.
RIGOL Sicherheits Begriffe und Symbole Begriffe in diesem Guide. Diese Begriffe können in diesem Handbuch auftauchen: WARNING Die Kennzeichnung WARNING beschreibt Gefahrenquellen die leibliche Schäden oder den Tod von Personen zur Folge haben können. CAUTION Die Kennzeichnung Caution (Vorsicht) beschreibt Gefahrenquellen die Schäden am Gerät hervorrufen können.
RIGOL Measurement Category MSO4000/DS4000 series digital oscilloscopes can make measurements in measurement category I. WARNING This oscilloscope can only be used for measurements within its specified measurement categories. Measurement Category Definitions Measurement category I is for measurements performed on circuits not directly connected to MAINS.
RIGOL Ventilation Requirement This oscilloscope uses fan to force cooling. Please make sure that the air intake and exhaust areas are free from obstructions and have free air. When using the oscilloscope in a bench-top or rack setting, provide at least 10 cm clearance beside, above and behind the instrument for adequate ventilation.
RIGOL Working Environment Temperature Operating: 0℃ to +50℃ Non-operating: -40℃ to +70℃ Humidity 0℃ to +30℃: ≤95% relative humidity +30℃ to +40℃: ≤75% relative humidity +40℃ to +50℃: ≤45% relative humidity WARNING To avoid short circuits or electric shock, please do not operate in humid environment.
Page 15
RIGOL Pollution Degree Degree 2 Pollution Degree Definitions Pollution degree 1: No pollution or only dry, non-conductive pollution occurs. The pollution has no influence. For example, a clean room or air-conditioned office environment. Pollution degree 2: Normally only dry, non-conductive pollution occurs. An occasional and temporary conductive environment may caused by condensation.
RIGOL General Care and Cleaning General Care Do not store or leave the instrument where it may be exposed to direct sunlight for long periods of time. Cleaning Clean the instrument regularly according to its operating conditions. To clean the exterior surface, perform the following steps: Disconnect the instrument from all power sources.
RIGOL Environmental Considerations The following symbol indicates that this product complies with the WEEE Directive 2002/96/EC. Product End-of-Life Handling The equipment may contain substances that could be harmful to the environment or human health. In order to avoid the release of such substances into the environment...
MSO4000/DS4000 Series Overview MSO4000/DS4000 is a multifunctional and high-performance digital oscilloscope developed using RIGOL Ultra Vision chipset. MSO4000 series is a mixed-signal oscilloscope which is optimized for the debugging in embedded design and test areas that require simultaneous measurement of both analog and digital signals. MSO4000 provides superb specifications, convenient functions, and easy-to-use design which help users fulfill their tasks (such as measurement and remote control) more quickly.
Page 19
RIGOL Precise delayed sweep function Built-in FFT function Built-in logic analyzer function (MSO4000 series) Pass/Fail test function Multiple waveform math operation functions Standard interfaces: dual USB HOST, USB DEVICE, LAN and GPIB (optional) Support USB storage device and PictBridge printer ...
RIGOL Document Overview Main Topics in this manual: Chapter 1 Quick Start Introduce the preparations before using the instrument and introduce the instrument. Chapter 2 Vertical System Setting Introduce the functions of the vertical system of the oscilloscope. Chapter 3 Horizontal System Setting Introduce the functions of the horizontal system of the oscilloscope.
Page 21
RIGOL Chapter 11 Waveform Recording Introduce how to record waveform as well as how to play back and analyze the recorded waveform. Chapter 12 Display Control Introduce how to control the display of the oscilloscope. Chapter 13 Store and Recall Introduce how to perform external/internal storage and recall.
Page 22
RIGOL Format Conventions in this Manual: 1. Key The key at the front panel is denoted by the format of “Button Name (Bold) + Text Box”. For example, Utility denotes the “Utility” key. 2. Menu Softkey The menu softkey is denoted by the format of “Menu Word (Bold) + Character Shading”.
Page 23
User Manuals of this Product: The main manuals of this product include the Quick Guide, User’s Guide, Programming Guide and etc. For the newest version of the desired manual, please download it from RIGOL official website (www.rigol.com). MSO4000/DS4000 User’s Guide 800.544.2843 www.calcert.com...
RIGOL Contents Contents Guaranty and Declaration ................. I Safety Requirement ................. II General Safety Summary................II Safety Terms and Symbols ................ V Allgemeine Sicherheits Informationen ............VI Sicherheits Begriffe und Symbole .............. IX Measurement Category ................X Ventilation Requirement ................XI Working Environment ................
Page 25
Contents RIGOL SINGLE ..................1-23 Multi-function Knob ................ 1-24 Navigation Knob ................1-24 Default ..................1-25 Print ..................... 1-25 Function Keys ................1-26 Record ..................1-27 LOGIC ANALYZER................1-28 User Interface ..................1-29 To Use the Security Lock ............... 1-36 To Use the Built-in Help System .............
Page 26
RIGOL Contents Chapter 4 Sample System Setting ............4-1 Sample Mode ..................4-2 Sample Rate ..................4-3 LA Sample Rate ..................4-5 To Specify the Acquisition Mode ............... 4-6 Normal ................... 4-6 Average ..................4-6 Peak Detect ..................4-8 High Resolution ................4-8 To Set the Memory Depth ...............
Page 27
Contents RIGOL Digital Filter..................6-9 Logic Operation ................6-11 Advanced Operation ............... 6-14 MATH Waveform Vertical Control ............. 6-19 To Make Auto Measurement ..............6-20 Quick Measurement after AUTO............6-21 To Make One-key Measurement ............6-22 To Set the Threshold ..............6-30 To Specify the Measurement Range ..........
Page 28
RIGOL Contents CAN Decoding (Option) ................. 8-26 FlexRay Decoding (Option) ..............8-31 Chapter 9 Reference Waveform ............9-1 To Turn the Reference Waveform Channel On/Off ........9-2 To Select the Current Reference Waveform Channel ........9-3 To Set the Reference Waveform Color ............9-3 To Select the Reference Waveform Source Channel ........
Page 29
Contents RIGOL To Specify the Screen Grid Type ............. 12-5 To Set the Grid Brightness ..............12-5 To Set the Menu Display Time ............... 12-6 Chapter 13 Store and Recall ............13-1 Storage System ..................13-2 Storage Type ..................13-3 To Perform Internal Storage and Recall ...........
Page 30
RIGOL Contents Chapter 15 Remote Control ............15-1 Remote Control via USB ................ 15-2 Remote Control via LAN ................ 15-6 Remote Control via GPIB ............... 15-9 Chapter 16 Troubleshooting ............16-1 Chapter 17 Specifications ............17-1 Chapter 18 Appendix ..............18-1 Appendix A: Accessories and Options .............
Chapter 1 Quick Start RIGOL Chapter 1 Quick Start This chapter introduces the notices when using the oscilloscope for the first time, the front panel, rear panel and user interface as well as the using method of the built-in help system.
RIGOL would not be responsible for free maintenance/rework or replacement of the unit. 2. Inspect the instrument In case of any damage, or defect, or failure, notify your RIGOL sales representative. 3. Check the accessories Please check the accessories according to the packing lists. If the accessories are incomplete or damaged, please contact your RIGOL sales representative.
RIGOL Chapter 1 Quick Start To Prepare the Oscilloscope for Use To Remove the Protective Cover Before using the oscilloscope, remove the front panel protective cover by grasping the transverse tabs on each side and pull them outward. Front Panel Protective Cover...
Chapter 1 Quick Start RIGOL To Adjust the Supporting Legs Adjust the supporting legs properly to use them as stands to tilt the oscilloscope upwards for stable placement of the oscilloscope as well as better operation and observation. Unfold or fold the supporting legs in the directions of the arrows as shown in the figures below.
RIGOL Chapter 1 Quick Start To Connect to Power Supply This oscilloscope accepts two kinds of AC power supplies: 100-127 V, 45-440 Hz and 100-240 V, 45-65 Hz. Please use the power cord supplied with the accessories to connect the oscilloscope to the AC power source via the power socket as shown in the figure below.
Chapter 1 Quick Start RIGOL Power-on Inspection When the oscilloscope is energized, press the power key at the lower-left side of the front panel to start the oscilloscope. During the start-up process, the oscilloscope performs a series of self-test items and you can hear the sounds of relay switching.
Chapter 1 Quick Start To Connect the Probe RIGOL provides passive and active probes for DS4000 as well as passive, active and logic probes for MSO4000. For detailed technical information of the probes, please refer to the corresponding Probe User’s Guide. The probes recommended for this oscilloscope are as shown in the table below.
Page 39
Chapter 1 Quick Start RIGOL To Connect the Passive Probe 1. Connect the BNC terminal of the probe to the analog channel input terminal or external trigger input connector of the oscilloscope at the front panel (please refer to “Front Panel Overview”).
Page 40
RIGOL Chapter 1 Quick Start To Connect the Active Probe Take RP7150 (with differential probe front) as an example. Connect the probe front with the active probe amplifier. Connect the other end of the active probe amplifier to the analog channel input terminal or external trigger input connector of the oscilloscope at the front panel (please refer to “Front Panel Overview”).
Page 41
To Connect the Logic Probe 1. Connect the single-wire end of the logic probe to the digital channel input terminal [LOGIC D0-D15] at the front panel of MSO4000 series digital oscilloscope in the correct direction. 2. Connect the other end of the logic probe to the signal under test. RPL2316...
Page 42
RIGOL Chapter 1 Quick Start To Connect the Tek Probe Connect the Tek probe to one end of T2R1000 adaptor correctly. Connect the other end of T2R1000 adaptor to the analog channel input terminal or external trigger input connector (refer to “Front Panel Overview”) of MSO4000/DS4000 series oscilloscope at the front panel correctly.
Chapter 1 Quick Start RIGOL Passive Probe Function Inspection&Compensation Adjustment After connecting the passive probe, you need to perform probe function inspection and probe compensation before making measurements. Function Inspection You need to perform probe function inspection according to the steps below before making measurements using the passive probe.
Page 44
RIGOL Chapter 1 Quick Start Figure 1-10 Probe Compensation Signal (Square Waveform) 6. Use the same method to test the other channels. WARNING To avoid electric shock, please make sure that the insulated wire of the probe is in good condition and do not touch the metallic part of the probe when the probe is connected to a high voltage source.
Page 45
Chapter 1 Quick Start RIGOL Probe Compensation When the probes are used for the first time or when the probe compensation signal does not match that shown in Figure 1-10 (refer to “Function Inspection”), you should compensate the probes to match the input channels of the oscilloscope.
RIGOL Chapter 1 Quick Start Front Panel Overview 9 10 Figure 1-12 Front Panel Overview Table 1-1 Front Panel Description Description Description 29 Parameters Measurement Menu Record Softkeys Power Key Multi-function Knob USB HOST Navigation Knob Digital Channel Input Terminal...
Chapter 1 Quick Start RIGOL Rear Panel Overview Figure 1-13 Rear Panel Overview 1. Trig Out/Calibration Various kinds of signals can be output from this connector (press Utility AuxOutput to select the desired output type). 1) TrigOut: the oscilloscope outputs a signal that can reflect the current capture rate of the instrument each time a trigger is generated.
Page 48
You can connect a USB storage device via this interface to perform external storage and recall. You can also extend a GPIB interface via this interface using the USB-GPIB interface converter (optional) provided by RIGOL to realize GPIB communication. Note: There is also a USB HOST interface at the front panel.
Chapter 1 Quick Start RIGOL Front Panel Function Overview VERTICAL MSO4000/DS4000 provides an independent vertical control system for each analog input channel. CH1, CH2, CH3, CH4: analog input channel switches. The four channels are marked by different colors which are also used to mark the corresponding waveforms on the screen and the channel input connectors.
RIGOL Chapter 1 Quick Start screen changes accordingly. Pressing down the knob can quickly switch the vertical scale adjustment mode between “Coarse” and “Fine”. MATH: press this key to open the math operation function menu. You can perform addition, subtraction, multiplication, division, FFT, digital filter, logic operations and advanced operations.
Page 51
Chapter 1 Quick Start RIGOL sweep time base. During the modification, the delayed sweep waveform is expanded or compressed horizontally; the delayed sweep time base message (as shown in the figure below) at the middle of the screen changes accordingly.
RIGOL Chapter 1 Quick Start TRIGGER MODE: press this key to switch the trigger mode to “Auto”, “Normal” or “Single” and the corresponding status backlight of the current trigger mode would be illuminated. TRIGGER LEVEL: trigger level knob. ...
Chapter 1 Quick Start RIGOL AUTO Press this key to enable the waveform auto setting function. The oscilloscope will automatically adjust the vertical scale, horizontal time base, trigger mode and so on according to the input signal to realize optimum waveform display. Besides, the quick parameter measurement function is also provided (for more details, refer to “Quick Measurement after...
RIGOL Chapter 1 Quick Start Multi-function Knob This knob can be used to select parameter item, adjust the value of the parameter and etc. In menu operation, this knob can be used to open the submenu or select the desired parameter item. For...
Chapter 1 Quick Start RIGOL Default Press this key and then press OK to restore the oscilloscope to its factory setting (please refer to Table 13-2). Print Press this key to execute the print function or save the screen content in a USB storage device as a picture file. If the...
RIGOL Chapter 1 Quick Start Function Keys Measure: press this key to enter the measurement setting menu. You can set the signal source, perform all measurement, turn on the statistic function, make measurement settings and so on. Acquire: press this key to enter the sample setting menu. You can set the ...
Chapter 1 Quick Start RIGOL Record Stop Play/Pause Record Record: press this key to start recording the waveform. The backlight is illuminated in red and flashes. Besides, the backlight will also be illuminated in red and flashes when the open mode (record constant on) is enabled until you press RUN/STOP to stop the waveform recording.
Chapter 1 Quick Start RIGOL User Interface The user interface of MSO4000/DS4000 series oscilloscope is as shown in Figure 1-14 and Table 1-2. Figure 1-14 User Interface MSO4000/DS4000 User’s Guide 1-29 800.544.2843 www.calcert.com sales@calcert.co...
Page 60
RIGOL Chapter 1 Quick Start Table 1-2 User Interface Overview Name Name Auto Measurement Items Trigger Setting Analog Channel Label and CH1 Status Label Waveform Digital Channel Label and CH2 Status Label Waveform Status CH3 Status Label Horizontal Time Base...
Page 61
The label and waveform of the digital channel currently selected are both displayed in red. Note: Only applicable to MSO4000 series oscilloscope. 4. Status Available statuses include RUN, STOP, T’D (triggered), WAIT and AUTO.
Page 62
RIGOL Chapter 1 Quick Start the screen. 9. Horizontal Position Use HORIZONTAL POSITION to modify this parameter. Pressing down the knob can quickly reset the horizontal position (zero). 10. Trigger Setting The trigger setting information includes trigger type, trigger source and trigger level/threshold level (may also include trigger level label).
Page 63
Chapter 1 Quick Start RIGOL 11. CH1 Status Label Display the on/off status of CH1. When it is on, the status label is highlighted. Display the vertical scale of CH1 (namely the voltage value per grid of the ...
Page 64
The digital channels turned off are displayed in gray. Note: This function is only applicable to MSO4000 series oscilloscope. 17. Notification Area Display the system time, sound icon, USB storage device icon and PictBridge printer icon.
Page 65
Chapter 1 Quick Start RIGOL PictBridge printer, is displayed. 18. Function Setting Menu Press MENU at the right of the screen continuously to unfold or fold the function setting menu. When the function setting menu is unfolded, press any softkey to activate the corresponding menu item (for menu items that include submenus or multiple parameter items, the submenus or parameter items will also be opened).
RIGOL Chapter 1 Quick Start To Use the Security Lock If needed, you can use a security lock (available separately) to lock the oscilloscope to a fixed location. The method is as follows. Align the lock with the lock hole and plug it straight into the lock hole, turn the key clockwise to lock the oscilloscope and then pull the key out.
Chapter 1 Quick Start RIGOL To Use the Built-in Help System The help system of this oscilloscope provides help information for all the keys and menu softkeys at the front panel. Press Help to open the help interface and press again to close the interface.
Page 68
RIGOL Chapter 1 Quick Start “To Index” and then press down the knob to switch to Index mode. Index: In this mode, you can turn the multi-function knob to select the desired item (the item currently selected is displayed with brown shading) in the “Help Options”...
Chapter 2 Vertical System Setting RIGOL Chapter 2 Vertical System Setting MSO4000/DS4000 provides four analog input channels (CH1 to CH4) and provides independent vertical control system for each channel. As the vertical system setting methods of the four channels are completely the same, this chapter takes CH1 as an example to illustrate the setting method of the vertical system.
RIGOL Chapter 2 Vertical System Setting To Turn the Analog Channel On/Off Connect a signal to the channel connector of CH1 and then press CH1 in the vertical control area (VERTICAL) at the front panel to turn CH1 on or off. When it is turned on, the backlight of CH1 is illuminated;...
Chapter 2 Vertical System Setting RIGOL To Adjust the Vertical Scale The vertical scale refers to the voltage value per grid on the vertical direction on the screen and is usually expressed as V/div. The adjustable range of the vertical scale is related to the current probe ratio and input impedance.
RIGOL Chapter 2 Vertical System Setting To Adjust the Vertical Position The vertical position refers to the position of the ground level of the waveform relative to the screen center in the vertical direction. The adjustable range of the vertical position is related to the current input impedance, probe ratio and vertical scale (refer to "Specifications").
Chapter 2 Vertical System Setting RIGOL To Specify the Channel Coupling Set the coupling mode to filter out the undesired signals. For example, the signal under test is a square waveform with DC offset. When the coupling mode is “DC”, the DC and AC components of the signal under ...
RIGOL Chapter 2 Vertical System Setting To Specify the Bandwidth Limit Setting the bandwidth limit can reduce display noise. For example, the signal under test is a pulse with high-frequency oscillations. When bandwidth limit is disabled, the high frequency components of the signal ...
To Set the Probe The analog channels of this oscilloscope support normal passive probes, active probes and Tek probes (with RIGOL T2R1000 active probe adaptor, option). The oscilloscope can automatically identify the type of the probe currently connected and the probe ratio. For the detailed technical information of the probes, please refer to the corresponding probe User’s Guide.
RIGOL Chapter 2 Vertical System Setting Passive Probe 1. Probe Type Probe Type displays “Nor-Probe”; for example, when RIGOL RP3500A probe is connected. 2. Ratio For some normal passive probes, the oscilloscope can identify the probe ratio automatically. If not, press Ratio and turn the multi-function knob to select the corresponding probe ratio;...
RIGOL Active Probe 1. Probe Type Probe Type displays “Active Probe”; for example, when RIGOL RP7150 probe is connected. Note: When a 50 Ω active probe is connected, the input impedance of the channel (refer to “To Specify the Input Impedance”) will be set to “50 Ω”...
Page 78
RIGOL Chapter 2 Vertical System Setting connectors according to the probe calibration prompt message. Adjust the distance between the RP7150 probe tips. Connect the positive probe tip to the signal at the moddile of the calibration board and connect the negative probe tip to both sides of the signal line.
Chapter 2 Vertical System Setting RIGOL Tek Probe 1. Probe Type Probe Type displays “Tek Probe”; for example, when Tek (Tektronix) P6245 probe is connected. Note: When a Tek probe is used, the oscilloscope sets the channel input impedance automatically according to the model of the probe actually connected.
RIGOL Chapter 2 Vertical System Setting 6. Bias Voltage Press Bias Voltage and turn the multi-function knob to adjust the DC bias voltage. Note: The adjustable range of the DC bias voltage is related to the probe model and probe ratio (only for models that supports multiple probe ratios). For the details, please refer to the technical specifications of the corresponding probe.
Chapter 2 Vertical System Setting RIGOL To Turn Waveform Invert On/Off Press CH1 and its backlight is illuminated; then, press Invert to turn on or off waveform invert (the default is off). When waveform invert is turned off, the waveform display is normal; when waveform invert is turned on, the waveform voltage values are inverted (as shown in the figures below).
RIGOL Chapter 2 Vertical System Setting To Specify the Amplitude Unit Select the amplitude display unit for the current channel. The available units are “W”, “A”, “V” and “U”. When the unit is changed, the amplitude unit displayed in the channel status label will change accordingly.
Chapter 2 Vertical System Setting RIGOL To Set the Channel Label By default, the oscilloscope uses the number of the channel (such as , displayed at the left of the screen) to mark the corresponding analog channel. For easier use, you can define a label for each analog channel.
Page 84
RIGOL Chapter 2 Vertical System Setting Press Keyboard to select the “Keyboard” area, turn the multi-function knob to select “C” and press the knob to input this character in the “Name Input Area”. Use the same method to input the remaining characters “hn1”. Pay attention to the cases of the letters during the input process.
Chapter 2 Vertical System Setting RIGOL To Set the Delay Calibration When using an oscilloscope for actual measurements, the transmission delay of the probe cable may bring greater error (zero offset). Zero offset is defined as the offset of the crossing point of the waveform and trigger level line relative to the actual trigger position, as shown in the figure below.
Chapter 3 Horizontal System Setting RIGOL Chapter 3 Horizontal System Setting The contents of this chapter: To Adjust the Horizontal Time Base To Adjust the Horizontal Position To Specify the Time Base Mode To Turn Delayed Sweep On/Off ...
RIGOL Chapter 3 Horizontal System Setting To Adjust the Horizontal Time Base The horizontal time base (also called horizontal scale) refers to the time value per grid in the horizontal direction on the screen and is usually expressed as s/div. The adjustable range of the horizontal time base is related to the instrument model, as shown in the table below.
Chapter 3 Horizontal System Setting RIGOL To Adjust the Horizontal Position The horizontal position (also called trigger position) refers to the position of the trigger point of the waveform relative to the screen center in the horizontal direction. The horizontal position is positive (negative) when the waveform trigger point is at the left (right) of the screen center.
RIGOL Chapter 3 Horizontal System Setting To Specify the Time Base Mode Press MENU in the horizontal control area (HORIZONTAL) at the front panel; then, press Time Base, turn the multi-function knob to select the desired time base mode and press the knob. You can also press Time Base continuously to switch the current time base mode.
Chapter 3 Horizontal System Setting RIGOL X-Y Mode In this mode, the X axis and Y axis both represents voltage. The oscilloscope automatically turns on all the four channels (CH1 to CH4) and the screen is divided into two cursor areas (XY1 and XY2). In XY1 area (at the left of the screen), the X axis and Y axis track the voltages of CH1 and CH2 respectively;...
Page 92
RIGOL Chapter 3 Horizontal System Setting Phase Deviation Measurement Principle The phase deviation between two signals with the same frequency can be easily measured via the Lissajous method. The figure below shows the measurement principle of the phase deviation (assume that the amplitudes of the two signals are the same).
Page 93
Chapter 3 Horizontal System Setting RIGOL Application example: measure the phase deviation of the input signals of two channels. Method 1: Use Lissajous method Connect a sine signal to CH1 and then connect a sine signal with the same frequency and amplitude but a 90° phase deviation to CH2.
Page 94
RIGOL Chapter 3 Horizontal System Setting Method 2: Use the one-key measurement function of the oscilloscope Please measure the phase deviation between the input waveforms of CH1 and CH2 using the one-key measurement function by referring to “To Make One-key Measurement”.
Chapter 3 Horizontal System Setting RIGOL Roll Mode In this mode, the waveform scrolls from the right to the left. The adjustable range of the horizontal time base is from 200.0 ms/div to 1.000 ks/div. The horizontal time base will be set to 200.0 ms/div automatically when you turn on the “Roll” mode.
RIGOL Chapter 3 Horizontal System Setting To Turn Delayed Sweep On/Off Delayed sweep can be used to enlarge a length of waveform horizontally to view the waveform details. Note: Delayed sweep can only be turned on when the current time base mode is “Y-T”...
Page 97
Chapter 3 Horizontal System Setting RIGOL Normal Sweep Waveform Main Time Base Delayed Sweep Horizontal Position The waveform before enlargement Delayed Sweep Time Base Delayed Sweep Waveform (the waveform after enlargement) The waveform before enlargement: The waveform of normal sweep is displayed in the upper part of the screen. Wherein,...
Page 98
RIGOL Chapter 3 Horizontal System Setting the delayed sweep time base improves the waveform resolution (as shown in the figure above). Note: The delayed sweep time base should be less than or equal to the main time base. 3-12 MSO4000/DS4000 User’s Guide 800.544.2843...
Chapter 3 Horizontal System Setting RIGOL To Specify the Horizontal Reference The horizontal reference is the reference position according to which the screen waveform is expanded or compressed horizontally when using HORIZONTAL SCALE to modify the horizontal time base. Note: You can only set the horizontal reference when the “Y-T” mode is used and the delayed sweep is on.
Chapter 4 Sample System Setting RIGOL Chapter 4 Sample System Setting The contents of this chapter: Sample Mode Sample Rate LA Sample Rate To Specify the Acquisition Mode To Set the Memory Depth To Turn Anti-aliasing On/Off ...
RIGOL Chapter 4 Sample System Setting Sample Mode This oscilloscope only supports real-time sampling. In this mode, the oscilloscope samples and displays waveforms within a trigger event. When the sample is stopped, the oscilloscope will hold the last display. At this point, you can still use the vertical control and horizontal control to pan and zoom the waveform.
Chapter 4 Sample System Setting RIGOL Sample Rate Sample refers to the process that the oscilloscope converts analog signals to digital signals at a certain time interval and stores the data in order. The sample rate is the reciprocal of the time interval.
Page 104
RIGOL Chapter 4 Sample System Setting 2. Waveform Confusion: when the sample rate is lower than twice the actual signal frequency (Nyquist Frequency), the frequency of the waveform rebuilt from the sample data is lower than the actual signal frequency.
Chapter 4 Sample System Setting RIGOL LA Sample Rate The LA sampling refers to the process that the oscilloscope samples the compared digital signals at a certain time interval. The LA sample rate is the reciprocal of the time interval. For example, the oscilloscope with 1 GHz LA sample rate samples digital signals every 1 ns.
RIGOL Chapter 4 Sample System Setting To Specify the Acquisition Mode The acquisition mode is used to control how to generate waveform points from sample points. Press Acquire Acquisition and turn the multi-function knob to select the desired acquisition mode; then, press the knob. The acquisiton modes available include normal, average, peak detect and high-resolution.
Page 107
Chapter 4 Sample System Setting RIGOL The waveform before average: The waveform after 256 averages: MSO4000/DS4000 User’s Guide 800.544.2843 www.calcert.com sales@calcert.co...
RIGOL Chapter 4 Sample System Setting Peak Detect In this mode, the oscilloscope acquires the maximum and minimum values of the signal within the sample interval to get the envelope of the signal or the narrow pulse that might be lost. In this mode, signal confusion can be prevented but the noise displayed would be larger.
Chapter 4 Sample System Setting RIGOL To Set the Memory Depth The memory depth refers to the number of waveform points that the oscilloscope can store in a single trigger sample and it reflects the storage ability of the acquisition memory. MSO4000/DS4000 provides up to 140 M points memory depth.
Page 110
RIGOL Chapter 4 Sample System Setting accordingly. Note : When two channels use the same sampler and only one of the two channels is turned on, the system is in the interweave state; when both of the two channels are turned on, the system is in the non-interweave state.
Chapter 4 Sample System Setting RIGOL To Turn Anti-aliasing On/Off At slower sweep speed, the sample rate is reduced and the possibility of aliasing can be minimized by using a dedicated display algorithm. Press Acquire and press Anti-Aliasing to turn the anti-aliasing function on or off (the default is off).
Chapter 5 Trigger RIGOL Chapter 5 Trigger To generate a trigger, you set a trigger condition according to the requirement; when a waveform in the waveform stream meets this condition, the oscilloscope captures this waveform as well as the neighbouring part immediately and then displays the captured waveforms on the screen.
Chapter 5 Trigger RIGOL Trigger Source Signals input from the analog channels (CH1 to CH4), the external trigger input terminal [EXT TRIG] connector and [LOGIC D0-D15] connector (only for MSO4000) as well as the AC Line can all be used as trigger source.
Page 116
RIGOL Chapter 5 Trigger Note: When different trigger type is selected, the signal sources (one or more of the signals mentioned above) available are different. MSO4000/DS4000 User’s Guide 800.544.2843 www.calcert.com sales@calcert.co...
Chapter 5 Trigger RIGOL Trigger Level/Threshold Level The adjustment of the trigger level/threshold level is related to the trigger source type. When the trigger source is any of CH1 to CH4, turn TRIGGER LEVEL to adjust the trigger level.
RIGOL Chapter 5 Trigger Trigger Mode MSO4000/DS4000 provides Auto, Normal and Single trigger modes and the default is Auto. Pressing MODE in the trigger control area (TRIGGER) at the front panel can quickly switch the current trigger mode. You can also press MENU (in the trigger control area (TRIGGER) at the front panel) ...
Page 119
Chapter 5 Trigger RIGOL Trigger and the Pre-trigger/Post-trigger Buffer The following is the schematic diagram of the acquisition memory. To easily understand the trigger event, the acquisition memory is divided into the pre-trigger buffer and post-trigger buffer. Trigger Event Pre-trigger Buffer...
RIGOL Chapter 5 Trigger Trigger Coupling The trigger coupling setting determines the type of signal components that will be transmitted to the trigger circuit. Press MENU (in the trigger control area (TRIGGER) at the front panel) Setting Coupling and turn the multi-function knob to select the desired coupling type (the default is DC).
Chapter 5 Trigger RIGOL Trigger Holdoff The trigger holdoff can be used to set the period of time that the oscilloscope waits (namely the holdoff time) before re-arming the trigger circuitry after a trigger. The oscilloscope will not trigger until the holdoff time expires. Trigger holdoff can be used to stably trigger complex repeat waveforms (there are multiple edges (or other events) within a single repeating period of the waveform, such as pulse series).
RIGOL Chapter 5 Trigger Noise Rejection Noise rejection reduces the possibility of noise triggering by increasing the trigger hysteresis. Press MENU (in the trigger control area (TRIGGER) at the front panel) Setting and press Noise Reject to turn the noise rejection on or off.
Chapter 5 Trigger RIGOL Edge Trigger In edge trigger, the oscilloscope triggers on the trigger level/threshold level of the specified edge of the input signal. Press MENU (in the trigger control area (TRIGGER) at the front panel) Type, turn the multi-function knob to select “Edge”...
Page 124
RIGOL Chapter 5 Trigger 2. Specify the edge type Press Slope and turn the multi-function knob to select the kind of edge of the input signal on which the oscilloscope triggers. You can also press Slope continuously to switch the current edge type. The edge type currently selected is displayed at the upper-right side of the screen.
Chapter 5 Trigger RIGOL Pulse Trigger In pulse trigger, the oscilloscope triggers on the positive or negative pulse with a specified width. For this oscilloscope, the positive pulse width is defined as the time difference between the two crossing points of the trigger level and positive pulse and the negative pulse width is defined as the time difference between the two crossing points of the trigger level and negative pulse, as shown in the figure below.
Page 126
RIGOL Chapter 5 Trigger source. Turn TRIGGER LEVEL to adjust the trigger level (when the trigger source is CH1 to CH4 or EXT) or threshold level (when the trigger source is D0 to D15) (please refer to “Trigger Level/Threshold Level”). The current trigger level/threshold level is displayed at the upper-right side of the screen.
Page 127
Chapter 5 Trigger RIGOL adjust the pulse width within a relatively larger range, this function can help you quickly finish the setting. When the pulse width condition is set to press Upper Limit and Lower Limit respectively and use the...
RIGOL Chapter 5 Trigger Runt Trigger Runt trigger is used to trigger pulses that pass one trigger level but fail to pass the other one. For this oscilloscope, positive runt pulses are defined as pulses that pass the lower limit of the trigger level but fail to pass the upper limit of the trigger level; negative runt pulses are defined as pulses that pass the upper limit of the trigger level but fail to pass the lower limit of the trigger level, as shown in the figure below.
Page 129
Chapter 5 Trigger RIGOL 2. Adjust the trigger level In runt trigger, you need to set the trigger level upper limit and trigger level lower limit which correspond to the trigger labels respectively. Select the trigger level to be adjusted ...
Page 130
RIGOL Chapter 5 Trigger 3. Specify the pulse polarity Press Polarity to select the desired pulse polarity. : positive polarity. The oscilloscope triggers on the positive runt pulse. : negative polarity. The oscilloscope triggers on the negative runt pulse.
Page 131
Chapter 5 Trigger RIGOL adjust the maximum runt pulse width within a relatively larger range, this function can help you quickly finish the setting. <>: the oscilloscope triggers when the width of the runt pulse is lower than the specified maximum runt pulse width (Upper Limit) and greater than the specified minimum runt pulse width (Lower Limit).
RIGOL Chapter 5 Trigger Nth Edge Trigger In Nth edge trigger, the oscilloscope triggers on the nth edge that appears after the specified idle time. Edge Type: RisingEdge Number of Edges: 2 For example, in the waveform as shown in the figure above, the idle time should fulfill the following formula if the oscilloscope needs to trigger on the second rising edge appears after the specified idle time.
Page 133
Chapter 5 Trigger RIGOL source is CH1 to CH4) or threshold level (when the trigger source is D0 to D15) (please refer to “Trigger Level/Threshold Level”). The current trigger level/threshold level is displayed at the upper-right side of the screen.
Page 134
RIGOL Chapter 5 Trigger trigger control area (TRIGGER) at the front panel turns on. 6. Turn the noise rejection on or off Press Setting to enter the trigger setting submenu. In this submenu, press Noise Reject to turn the noise rejection on or off (please refer to “Noise Rejection”).
Chapter 5 Trigger RIGOL Slope Trigger In slope trigger, the oscilloscope triggers on the specified positive slope or negative slope with the specified time. For this oscilloscope, the positive slope time is defined as the time difference between the two crossing points of trigger level lines (the trigger level upper limit...
Page 136
RIGOL Chapter 5 Trigger 2. Adjust the trigger level In slope trigger, you need to set the trigger level upper limit and trigger level lower limit which correspond to the trigger labels respectively. Select the trigger level to be adjusted ...
Page 137
Chapter 5 Trigger RIGOL The calculation formula of the slope is: UpLevel - LowLevel SlewRate = Time Wherein, Time is the time setting value. ― When the slope condition is , the trigger level and slope range information is as shown in the figure below.
Page 138
RIGOL Chapter 5 Trigger than the specified lower limit of time and lower than the specified upper limit of time. 4. Set the slope time The slope time range is related to the current slope condition setting. When the slope condition is set to ...
Page 139
Chapter 5 Trigger RIGOL 6. Set the trigger holdoff and noise rejection Press Setting to enter the trigger setting submenu. In this submenu: 1) Press Holdoff to adjust the holdoff time (please refer to “Trigger Holdoff”). The range is from 100 ns to 10 s.
RIGOL Chapter 5 Trigger Video Trigger Video trigger is used to capture the complex waveforms of standard video signal. The trigger circuit checks the horizontal and vertical intervals of the waveform and triggers according to the video trigger setting. MSO4000/DS4000 supports the following video standards.
Page 141
Chapter 5 Trigger RIGOL 3. Specify the sync trigger mode Press Sync, turn the multi-function knob to select the desired sync trigger mode and press the knob. You can also press Sync continuously to switch the current sync trigger mode.
Page 142
RIGOL Chapter 5 Trigger video standard. You can also press Standard continuously to switch the current video standard. NTSC: the field frequency is 60 fields per second and the frame frequency is 30 frames per second. The TV scan line is 525 with the even field goes first and the odd field follows behind.
Chapter 5 Trigger RIGOL Pattern Trigger This trigger mode identifies a trigger condition by looking for a specified pattern. The pattern is a logical “AND” combination of the channels. Each channel can have a value of high (H), low (L) or don’t care (X). A rising or falling edge can be specified for one channel included in the pattern.
Page 144
RIGOL Chapter 5 Trigger selected is highlighted in red in the pattern setting area at the bottom of the screen. Note: Channels that are turned off can also be selected as the trigger source. 2) Specify the pattern of the current channel...
Page 145
Chapter 5 Trigger RIGOL 2. Specify the trigger mode Press Sweep to select “Auto”, “Normal” or “Single” (please refer to “Trigger Mode”). The corresponding status light of the trigger mode currently selected in the trigger control area (TRIGGER) at the front panel turns on.
RIGOL Chapter 5 Trigger RS232 Trigger MSO4000/DS4000 can trigger on the start of a frame, error frame, check error or the specified data bit of the RS232 signal. RS232 bus is a serial communication mode used for the data transmission between PCs or between PC and terminal end.
Page 147
Chapter 5 Trigger RIGOL to “Trigger Level/Threshold Level”). The current trigger level/threshold level is displayed at the upper-right side of the screen. 2. Specify the bus polarity Press Polarity to set the polarity of data transmission to “Normal” or “Invert”.
Page 148
RIGOL Chapter 5 Trigger The range of data is related to the “Data Bits” setting. When the data bits is set to 5 bit, 6 bit, 7 bit and 8 bit, the ranges of data are 1 to 31, 0 to 63, 0 to 127 and 0 to 255 respectively.
Page 149
Chapter 5 Trigger RIGOL 6. Turn the noise rejection on or off Press Setting to enter the trigger setting submenu. In this submenu, press Noise Reject to turn the noise rejection on or off (please refer to “Noise Rejection”). Note: When the trigger source is D0 to D15, the trigger coupling and noise rejection are invalid.
RIGOL Chapter 5 Trigger I2C Trigger MSO4000/DS4000 can trigger on the start condition, restart, stop, missing acknowledgement or on the read/write frame with specific device address and data value. The I2C bus is a two-wire serial bus consisting of the SCL and SDA. The SCL determines the transimission rate and the SDA determines the transmission data, as shown in the figure below.
Page 151
Chapter 5 Trigger RIGOL Note: Channels that are turned off can also be selected as the signal source of the SCL and SDA. Turn TRIGGER LEVEL to adjust the trigger level (when CH1 to CH4 is selected) or threshold level (when D0 to D15 is selected) (please refer to “Trigger Level/Threshold Level”).
Page 152
RIGOL Chapter 5 Trigger press Direction and turn the multi-function knob ― to select “Read”, “Write” or “R/W”. You can also press Direction continuously to switch the current data direction. Note: When the address bits is set to "8 bit", you do not need to set the data direction.
Page 153
Chapter 5 Trigger RIGOL displayed at the bottom of the screen, as shown in the figure below. The bit currently selected is highlighted in red. When the data parameter is modified, the data information changes accordingly. Please set the address and data parameters by referring to the introductions in "Address"...
RIGOL Chapter 5 Trigger SPI Trigger In SPI trigger, if the trigger condition (CS condition or timeout condition) is met, the oscilloscope will trigger when the specified data is found. The sequence chart of SPI bus is as shown in the figure below. SCL is the serial clock line, SDA is the serial data line and CS is the chip select line.
Page 155
Chapter 5 Trigger RIGOL Note: Channels that are turned off can also be selected as the signal source of the clock line and data line. Turn TRIGGER LEVEL to adjust the trigger level (when CH1 to CH4 is selected) or threshold level (when D0 to D15 is selected) (please refer to “Trigger Level/Threshold Level”).
Page 156
RIGOL Chapter 5 Trigger currently selected is displayed at the upper-right side of the screen. Note: Channels that are turned off can also be selected as the signal source of the CS line. Press Mode to set the CS mode to “...
Chapter 5 Trigger RIGOL CAN Trigger MSO4000/DS4000 can trigger on the start of a frame, end of a frame, frame of the specified type or error frame of the specified type of the CAN signal. The data frame format of the CAN bus is as shown in the figure below.
Page 158
RIGOL Chapter 5 Trigger 2. Specify the signal type Press Signal Type and turn the multi-function knob to select the desired signal type. You can also press Signal Type continuously to switch the current signal type. Rx: the receiving signal from the CAN signal line.
Page 159
Chapter 5 Trigger RIGOL You can make the following settings: Press ID Setup to select “Random ID” or “Specific ID”. When “Random ID” is selected, the ID data will not be used as the trigger condition. When “Specific ID” is selected, press ID Format to select “Standard”...
Page 160
RIGOL Chapter 5 Trigger also press Error Type continuously to switch the current error type. The error types available include “Bit Fill”, “Answer Error”, “Check Error”, “Format Error” and “Random Error”. 4. Set the signal rate Press Baud and turn the multi-function knob to select a CAN signal rate that matches the CAN bus signal.
Page 161
Chapter 5 Trigger RIGOL position with a step of 1%. The range is from 5% to 95%. 6. Specify the trigger mode Press Sweep to select “Auto”, “Normal” or “Single” (please refer to “Trigger Mode”). The corresponding status light of the trigger mode currently selected in the trigger control area (TRIGGER) at the front panel turns on.
RIGOL Chapter 5 Trigger FlexRay Trigger MSO4000/DS4000 can trigger on the specified frame, symbol, error or TSS of the FlexRay bus. FlexRay is a type of differential serial bus configured with three continuous segments (namely packet header, payload and packet trailer). Its data transmission rate is up to 10 Mbps.
Page 163
Chapter 5 Trigger RIGOL Note: Channels that are turned off can also be selected as the trigger source. Turn TRIGGER LEVEL to adjust the trigger level (when the trigger source is CH1 to CH4) or threshold level (when the trigger source is D0 to D15) (please refer to “Trigger Level/Threshold Level”).
Page 164
RIGOL Chapter 5 Trigger the trigger control area (TRIGGER) at the front panel turns on. 5. Turn the noise rejection on or off Press Setting to enter the trigger setting submenu. In this submenu, press Noise Reject to turn the noise rejection on or off (please refer to “Noise Rejection”).
Chapter 5 Trigger RIGOL USB Trigger MSO4000/DS4000 can trigger on a Start of Packet (SOP), End of Packet (EOP), Reset Complete (RC), Enter Suspend (Suspend) or Exit Suspend (ExitSuspend) of the differential USB data lines (D+ and D-). This trigger supports USB low speed and full speed. The figure below shows the USB data transmission protocol.
Page 166
RIGOL Chapter 5 Trigger source is CH1 to CH4) or threshold level (when the trigger source is D0 to D15) (please refer to “Trigger Level/Threshold Level”). The current trigger level/threshold level is displayed at the upper-right side of the screen.
Chapter 6 Operations and Measurements RIGOL Chapter 6 Operations and Measurements MSO4000/DS4000 can perform math operations, auto measurements and cursor measurements on sampled and displayed data. The contents of this chapter: To Make Math Operation To Make Auto Measurement ...
RIGOL Chapter 6 Operations and Measurements To Make Math Operation MSO4000/DS4000 can realize multiple kinds of math operations between waveform of different channels, including the basic operations (addition, subtraction, multiplication and division), FFT, digital filter, logic operations and advanced operations. The results of math operations also allow for further measurements (for details, please refer to “To Make Cursor Measurement”).
Chapter 6 Operations and Measurements RIGOL Basic Operations MSO4000/DS4000 supports the basic operations (addition, subtraction, multiplication and division) on the analog channels. You can perform the basic operations following the steps below. 1. Specify the basic operation Press MATH Operate, turn the multi-function knob...
Page 170
RIGOL Chapter 6 Operations and Measurements respectively. For the adjustment methods, please refer to “To Adjust the Horizontal Time Base” and “To Adjust the Horizontal Position”. Note: Turning HORIZONTAL SCALE and HORIZONTAL POSITION will adjust the horizontal scales and horizontal positions of the analog channel waveforms and math operation waveform at the same time.
Chapter 6 Operations and Measurements RIGOL The FFT (Fast Fourier Transform) operation can transform time domain signals to frequency domain signals (frequency spectrum). FFT operation can facilitate the following works: Measure the harmonic components and distortion in the system ...
Page 172
RIGOL Chapter 6 Operations and Measurements 2. Specify the FFT source Press Source, turn the multi-function knob to select the desired channel and press the knob. You can also press Source continuously to switch the channel currently selected. The channels available are CH1, CH2, CH3 and CH4.
Page 173
Chapter 6 Operations and Measurements RIGOL Table 6-1 Window Functions Waveforms Suitable for Window Characteristics Measurement Transient or short pulse, the signal The best frequency levels before and after the resolution. multiplication are basically the same. The poorest amplitude Sine waveforms with the same Rectangle resolution.
Page 174
RIGOL Chapter 6 Operations and Measurements dB: use the logarithmic mode to display the vertical amplitude. Vrms: use the linear mode to display the vertical amplitude. Note: If you need to display the FFT frequency spectrum in a relatively larger dynamic range, dB is recommended.
Chapter 6 Operations and Measurements RIGOL Digital Filter MSO4000/DS4000 provides 4 types of digital filters (Low Pass Filter, High Pass Filter, Band Pass Filter and Band Reject Filter). The frequencies within the specified waveform segment of the signal can be filtered by setting the range of bandwidth.
Page 176
RIGOL Chapter 6 Operations and Measurements Press Upper Limit and turn the multifunction knob to adjust the frequency upper limit of the filter. Press Lower Limit and turn the multifunction knob to adjust the frequency lower limit of the filter.
Chapter 6 Operations and Measurements RIGOL Logic Operation In logic operation, the oscilloscope performs logic operations on the logic levels of the channel waveforms of source A and source B point by point and display the results. When the voltage value of the source channel waveform is greater than the corresponding threshold, it is regarded as logic “1”;...
Page 178
RIGOL Chapter 6 Operations and Measurements You can perform the logic operations following the steps below. 1. Select the logic operation Press MATH Operate, turn the multi-function knob to select “Logic” and press the knob. You can also press Operate continuously to switch to the “Logic”...
Page 179
Chapter 6 Operations and Measurements RIGOL 6. Ajust the thresholds of source A and source B When source A or source B is set to an analog channel (CH1 to CH4), you can set the threshold of the corresponding source channel. When the voltage value of the source channel waveform is greater than the corresponding threshold, it is regarded as logic “1”;...
RIGOL Chapter 6 Operations and Measurements Advanced Operation The advanced operation function allows users to define operation functions. The oscilloscope will perform operations according to the expression defined by users and display the results. You can perform advanced operations following the steps below.
Page 181
Chapter 6 Operations and Measurements RIGOL 6. Turn the invert on or off Press Invert to turn the inverted display of the math operation waveform on or off. 7. Define the math operation waveform label Press Label to open the label submenu and the label modification interface.
Page 182
RIGOL Chapter 6 Operations and Measurements Function The function of each function is as shown in the table below. Note: The left brackets “(” here are only used to facilitate your entry and they are not a part of the function names.
Page 183
Chapter 6 Operations and Measurements RIGOL Operator The function of each operator is as shown in the table below. Table 6-4 Operator Operator Function + - * / Arithmetic operators: add, subtract, multiply and divide Parentheses, used to increase the priority of the operation...
Page 184
RIGOL Chapter 6 Operations and Measurements (Expression will automatically switch to “OFF”). At this point, the current expression is displayed at the bottom of the screen. The oscilloscope will operate according to this expression and display the results. To Set the Variables You can specify the variable by setting the “Mantissa”...
Chapter 6 Operations and Measurements RIGOL MATH Waveform Vertical Control Press MATH Operate and select the desired math operation to open the corresponding math operation submenu. In the submenu, you can adjust the vertical scale and vertical position of the math operation waveform according to your need.
RIGOL Chapter 6 Operations and Measurements To Make Auto Measurement MSO4000/DS4000 provides auto measurements of 29 waveform parameters and the statistics and analysis of the measurement results. What’s more, you can also use the frequency counter to realize more precise frequency measurement.
Chapter 6 Operations and Measurements RIGOL Quick Measurement after AUTO When a valid signal is input into the oscilloscop, press AUTO to enable the waveform auto setting function and open the auto setting menu. In this menu, pressing the specified menu the oscilloscope will execute the corresponding operation and display the corresponding waveform and measurement results.
RIGOL Chapter 6 Operations and Measurements To Make One-key Measurement MSO4000/DS4000 provides the one-key measurement function through which you can measure the 29 kinds of waveform parameters as shown in the table below. Please turn on the channel to be measured before making measurement.
Page 189
Chapter 6 Operations and Measurements RIGOL the measurement source by switching the current active channel (press CH1 to CH4 at the front panel). ― For the digital channels (D0 to D15), the “One-key measurement” function only supports period, frequency, +Width, -Width, +Duty Cycle and –Duty Cycle measurements.
Page 190
RIGOL Chapter 6 Operations and Measurements into the current measurement source or the measurement result is out of the valid range (too large or too small). Note: Modifying the threshold or measurement range might modify the measurement result. You can acquire the desired measurement result by “To Set the Threshold”...
Page 191
Chapter 6 Operations and Measurements RIGOL Time Parameters Fall Time Rise Time Threshold Upper Limit +Width -Width Threshold Middle Value Threshold Lower Limit Period 1. Period: the time between the threshold middle value of one edge to the threshold middle value of the next edge with the same polarity.
Page 192
RIGOL Chapter 6 Operations and Measurements Delay and Phase Parameters Period Source A Delay Source B 1. Dly A →B : the time difference (delay) between the threshold middle values of the rising edges of source A and source B. Negative delay indicates that the rising edge of source A occurred after the rising edge of source B.
Page 193
Chapter 6 Operations and Measurements RIGOL operation formula is: DelayA PhaseA × ° Period sourceA Wherein, denotes phase A →B ; denotes delay PhaseA DelayA A →B ; denotes the period of source A. Period sourceA Phs A →B : the phase difference (expressed in degree) between the threshold middle value of the rising edge of source A and the threshold middle value of the falling edge of source B.
Page 194
RIGOL Chapter 6 Operations and Measurements Voltage Parameters Overshoot Vmax Vtop Vamp Vbase Vmin Preshoot 1. Vmax: the voltage value from the highest point of the waveform to the GND. 2. Vmin: the voltage value from the lowest point of the waveform to the GND.
Page 195
Chapter 6 Operations and Measurements RIGOL the number of points being measured. 9. Vrms-1: the root mean square value of the voltage values of the waveform points within a single period. The operation formula is as shown in the formula above.
RIGOL Chapter 6 Operations and Measurements To Set the Threshold The threshold setting is used to set the threshold upper limit, middle value and lower limit (expressed as the percentages of the vertical amplitude) when performing auto measurement on the analog channels. Modifing the threshold values will affect the measurement results of all the time, delay and phase parameters.
Chapter 6 Operations and Measurements RIGOL To Specify the Measurement Range Press Measure Range to open the measurement range setting menu. In this menu, press Range to set the measurement range of one-key measurement to “Screen” or “Cursor”. When “Cursor” is selected, two cursor lines (as shown in the figure below) appear on the screen.
RIGOL Chapter 6 Operations and Measurements To Clear the Measurement Results The oscilloscope allows users to clear or recover the measurement results of one-key measurement. If you have currently enabled one or more items of the 29 measurement parameters, you can clear or recover the measurement results of at most 5 measurement items that are enabled lastly.
Chapter 6 Operations and Measurements RIGOL To Turn the Statistic Function On/Off The oscilloscope provides the statistic function of the measurement results of one-key measurement. You can make statistics of at most 5 measurement items that are enabled lastly and display the statistic results.
RIGOL Chapter 6 Operations and Measurements To Turn the Measurement History On/Off This oscilloscope allows users to view the measurement history (only the measurement items of which the measurement results are currently displayed) of one-key measurement. Press Measure History to open the measurement history menu. In this menu, press History to turn on or off the measurement history.
Chapter 6 Operations and Measurements RIGOL To Make All Measurement The all measurement function can measure all the time, voltage and area parameters (21 items in total, please refer to Table 6-5) of the waveform(s) of the specified channel(s) (one or more of the waveforms of CH1 to CH4 and MATH) and display the measurement results on the screen.
RIGOL Chapter 6 Operations and Measurements To Turn the Frequency Counter On/Off The hardware frequency counter supplied with this oscilloscope can make more precise measurement of the input signal frequency. Press Measure Counter and turn the multi-function knob to select the desired channel as the measurement source of the frequency counter.
Chapter 6 Operations and Measurements RIGOL To Make Cursor Measurement Cursors are the horizontal and vertical marks that can be used to measure the X axis values and Y axis values on a selected waveform. All the 29 waveform parameters (please refer to Table 6-5) supported by one-key measurement can be measured through cursor measurement.
Page 204
RIGOL Chapter 6 Operations and Measurements then press the knob. You can also press Cursor or Mode continuously to switch the current cursor mode. The modes available are “Manual”, “Track”, “Auto” and “X-Y”. When “OFF” is selected, the cursor measurement function is turned off.
Chapter 6 Operations and Measurements RIGOL Manual Mode In this mode, you can adjust the cursors manually to measure the values at the cursors of the specified source channel waveform. The measurement results will be displayed at the upper-left side of the screen. When the measurement parameters are changed, the measurement results will change accordingly.
Page 206
RIGOL Chapter 6 Operations and Measurements You can use cursor to perform manual measurement on the waveform of CH1 to CH4 or MATH channel following the steps below. 1. Turn on the manual measurement function Press Cursor Mode, turn the multi-function knob to select “Manual”...
Page 207
Chapter 6 Operations and Measurements RIGOL ― You can select measurement items of the X cursors and Y curosrs by setting the “horizontal measurement unit” (Time Unit) and The measurement results include A->X, B->X, A->Y, B->Y, △X, 1/△X “vertical measurement unit” (Vertical Unit) respectively.
Page 208
RIGOL Chapter 6 Operations and Measurements △X and 1/△X. Wherein, A->X, B->X and △X denotes frequency and position as reference). The measurement results include A->X, B->X, 1/△X denotes time. results include A->X, B->X and △X (expressed in degree). °: measure the phase values at the X cursors. The measurement ...
Page 209
Chapter 6 Operations and Measurements RIGOL B->Y and △Y (the unit is the same as that of the current source). include A->Y, B->Y and △Y (expressed in percentage). %: measure the ratio values at the Y cursors. The measurement results ...
Page 210
RIGOL Chapter 6 Operations and Measurements Press CursorAB and turn the multi-function knob to adjust the vertical positions of cursors A and B (Y cursors) at the same time. The adjustable range is limited within the screen. The vertical distance between cursor A and B (Y cursors) keeps unchanged.
Page 211
Chapter 6 Operations and Measurements RIGOL When using cursor to perform manual measurement on the waveform of the LA channel, the display mode is fixed at “X” and cannot be modified (only the X cursors are displayed). When different horizontal measurement units are selected, the cursor measurement results are different.
Page 212
RIGOL Chapter 6 Operations and Measurements 2. Specify the measurement source Press Source and turn the multi-function knob to select “LA”. You can also press Source continuously to switch the current source to “LA”. If “None” is selected, no cursor and measurement result will be displayed.
Page 213
Chapter 6 Operations and Measurements RIGOL cursor A and cursor B positions are defined to “0°” and “360°” phase positions respectively; at the same time, two unmovable vertical cursor lines (with relatively lower brightness) are displayed on the screen as the phase reference positions (“0°”...
RIGOL Chapter 6 Operations and Measurements Track Mode In this mode, when adjusting the X cusors horizontally, the Y cursors will automatically track the vertical amplitude at the X cursors of the correspionding source channel waveform. The crossing points of X cursors and Y cursors are marked by an orange rectangle (cursor A) and rhombus (cursor B).
Page 215
Chapter 6 Operations and Measurements RIGOL You can use cursor to perform track measurement on the waveform of CH1 to CH4 or MATH channel following the steps below. 1. Turn on the track measurement function Press Cursor Mode, turn the multi-function knob to select “Track”...
Page 216
RIGOL Chapter 6 Operations and Measurements The adjustable range is limited within the screen. The two crossing points are marked by an orange rectangle (cursor A) and an orange rhombus (cursor B) respectively. Note: When any of CursorA, CursorB and CursorAB is currently selected,...
Chapter 6 Operations and Measurements RIGOL Auto Mode You can use cursor to perform auto measurement on any of the 29 waveform parameters. In this mode, one or more cursors will appear. The system will automatically adjust the cursor positions as well as measure and calculate the corresponding parameters according to the signal variation.
Page 218
RIGOL Chapter 6 Operations and Measurements Meas.Para. will change accordingly. Measurement Example: Measure the delay between the rising edges of CH1 and CH2 waveforms using “Auto Measurement”. 6-52 MSO4000/DS4000 User’s Guide 800.544.2843 www.calcert.com sales@calcert.co...
Chapter 6 Operations and Measurements RIGOL X-Y Mode In this mode, the X cursors and Y cursors denote the waveform amplitudes of the specified channels respectively. You can adjust the cursor positions to measure the waveform amplitudes at the crossing points of cursors.
Page 220
RIGOL Chapter 6 Operations and Measurements respectively. XY2: the X cursors and Y cursors are used to measure the amplitude values of the specified points (cursor crossing points) of CH3 and CH4 waveforms respectively. 3. Ajust the X/Y cursor positions 1) Ajust the X cursor positions Press Type to select “X”.
RIGOL Chapter 7 Logic Analyzer MSO4000 series mixed signal digital oscilloscope provides the logic analyzer (LA) function. The logic analyzer consists of 16 logic channels (namely, digital channels; the default channel labels are D15 to D0). The oscilloscope compares the voltage values obtained in each sample of the digital channel with the preset logic threshold.
RIGOL Chapter 7 Logic Analyzer To Turn the Digital Channels On/Off You can turn on or off all the digital channels (D0 to D15) according to your needs. The channels are independent from one another. Press LA On/Off to enter the “On/Off” setting menu of the digital channels. You can set the on/off states of the digital channels using any of the following methods via this menu.
Chapter 7 Logic Analyzer RIGOL To Arrange the Digital Channels You can set the screen display positions of the waveforms of the digital channels currently turned on according to your needs. In addition, you can also quickly restore the sequence of all the digital channels on the screen using the auto arrangement operation.
RIGOL Chapter 7 Logic Analyzer Note: For the waveform and label colors of the active channel(s) and other channels currently turned on, please refer to “Digital Channel Waveform and Label Colors”. Auto Arrangement Press LA and then press ReOrder to “D0-D15” or “D15-D0” (the default is “D0-D15”).
Chapter 7 Logic Analyzer RIGOL 2. Set the digital channels included in Group1 Add digital channels Turn the multi-function knob to select any channel (the selecting state ); press the knob or press Group1. At this point, the selecting...
RIGOL Chapter 7 Logic Analyzer Note: “L” is only available when the number of digital channels currently turned on is not more than 8. If the current display size of the digital channel waveforms is “L”, WaveSize will ...
Chapter 7 Logic Analyzer RIGOL reset the threshold level of channels D15 to D8 to 0 V). The adjustable range is from -20.0 V to +20.0 V. Note: The threshold levels of channels D7 to D0 and channels D15 to D8 are independent.
Page 228
RIGOL Chapter 7 Logic Analyzer 4. Specify the endian Press Endian to select “Normal” (D0 corresponds to the lowest bit of the bus data) or “Invert” (D0 corresponds to the highest bit of the bus data). The default is “Normal”.
Page 229
Chapter 7 Logic Analyzer RIGOL the jitter time when no reference clock is set for the digital bus. Press JitterReject to turn the jitter rejection function on or off. Press JitterTime and turn the multi-function knob or the inner ...
RIGOL Chapter 7 Logic Analyzer To Set the Digital Channel Label By default, the instrument uses D0 to D15 as the channel labels of the 16 digital channels. To identify different channels conveniently, you can define a label for each digital channel.
Chapter 7 Logic Analyzer RIGOL To Set the Delay Calibration of the Digital Channels When using an oscilloscope for actual measurement, the transmission delay of the probe cable may bring greater error (zero offset). Zero offset is defined as the offset of the crossing point of the waveform and threshold level line relative to the trigger position (refer to "To Set the Delay Calibration").
RIGOL Chapter 7 Logic Analyzer Digital Channel Waveform and Label Colors When the digital channel is turned on, the corresponding channel waveform and label are displayed on the screen. The channel waveform and label colors are related to the current active digital channel.
Chapter 8 Protocol Decoding RIGOL Chapter 8 Protocol Decoding Protocol analysis can be used to discover errors, debug hardware and accelerate development more conveniently; thus, guaranteeing fast and high-quality project execution. Protocol decoding is the basis of protocol analysis. Only protocol analysis based on correct decoding are reliable and correct decoding can provide more error information as well.
RIGOL Chapter 8 Protocol Decoding Parallel Decoding The parallel bus consists of a clock line and data lines. The oscilloscope will sample the channel data on the rising edge, falling edge or the rising&falling edges of the clock signal; then, combine the sample data as a parallel multi-channel bus and display the data as a single bus value.
Page 235
Chapter 8 Protocol Decoding RIGOL 2) Specify the edge type of sampling Press Slope and turn the multi-function knob to set the oscilloscope to sample the channel data on the rising edge ( ), falling edge ( ) or ) of the clock signal. You can also press Slope rising&falling edges (...
Page 236
RIGOL Chapter 8 Protocol Decoding Press Threshold to open the threshold setting submenu. In this submenu, 1) Select the specified analog channel Press Channel and turn the multi-function knob to select the analog channel (CH1 to CH4) that needs to specify the threshold. You can also press Channel continuously to switch the channel currently selected.
Page 237
Chapter 8 Protocol Decoding RIGOL of the bus. Pressing the knob can quickly reset the vertical offset (zero). 7. View the event table The event table displays the detailed information of the decoding in table form to view relatively longer decoded data.
Page 238
RIGOL Chapter 8 Protocol Decoding Note: The display format of the decoded data in the event table is determined by the bus display format currently selected (refer to “Format”). When adjusting the horizontal time base, the waveform displayed on ...
Chapter 8 Protocol Decoding RIGOL RS232 Decoding (Option) The RS232 serial bus consists of the transmitting data line (TX) and the receiving data line (RX). Its frame format is as shown in Figure 5-2. In RS232 decoding, you need to set the start bit, data bit width, check bit and stop bit for each frame of data.
Page 240
RIGOL Chapter 8 Protocol Decoding When the current trigger type is “RS232”, performing the copy operation at any time can quickly apply the current RS232 trigger setting (the specified parameters) to RS232 decoding. Then, you can still set the RS232 decoding parameters according to the introductions below.
Page 241
Chapter 8 Protocol Decoding RIGOL and 1 Mbps. When “User” is selected, press Setup and turn the multi-function knob or the inner knob of the navigation knob to adjust the baud rate at a relatively smaller step or turn the outer knob of the navigation knob to adjust the baud rate at a relatively larger step.
Page 242
RIGOL Chapter 8 Protocol Decoding determined by the polarity currently selected. The setting methods of the other parameters are as follows. Data Bits Press Data Bits and turn the multi-function knob to set the number of bits of each frame of data. It can be set to 5, 6, 7, 8 or 9 and the default is Check Bit ...
Page 243
Chapter 8 Protocol Decoding RIGOL the bus data to Hex, Decimal, Binary or ASCII. You can also press Format continuously to switch the current display format. Note: When the event table is turned on, the decoding data in the event table will also be displayed in the format selected here.
Page 244
RIGOL Chapter 8 Protocol Decoding The event table displays the detailed decoding information (include the decoded data (TX and/or RX) as well as the corresponding line number, time and error information (TX and/or RX)) of the waveform on the screen in time order.
Page 245
Chapter 8 Protocol Decoding RIGOL (00000100, LSB) The check bit detected is 1 After the decoder is set to none check, the decoding shows no error. End of Frame Error This error is generated when the end condition of the frame is not met. For example, when the Stop Bit is set to 1.5, an end of frame error as shown in...
Page 246
RIGOL Chapter 8 Protocol Decoding Note: When end of frame error and check error occur at the same time, two error marks will be displayed. 8-14 MSO4000/DS4000 User’s Guide 800.544.2843 www.calcert.com sales@calcert.co...
Chapter 8 Protocol Decoding RIGOL I2C Decoding (Option) I2C serial bus consists of the clock line (SCLK) and the data line (SDA). The oscilloscope samples the channel data on the rising edge of the clock signal (if the source channel is an analog channel, the system will judge each data point as logic “1”...
Page 248
RIGOL Chapter 8 Protocol Decoding 3. Set the Clock Signal (SCLK) and Data Signal (SDA) Set the SCLK Press SCLK; turn the multi-function knob to select the desired clock channel and press the knob. You can also press SCLK continuously to switch the current clock channel.
Page 249
Chapter 8 Protocol Decoding RIGOL Press Include R/W; when “Open” is selected, the "R/W" bit will be included in the address information and be at the lowest bit; when “Close” is selected, the "R/W" bit will not be included in the address information.
Page 250
RIGOL Chapter 8 Protocol Decoding information. The line currently selected is highlighted with brown shading and the decoded data of this line is dispalyed at the lower-left side of the event table. Note: The display format of the decoded data in the event table is determined ...
Page 251
Chapter 8 Protocol Decoding RIGOL Address Information (Write Address) 9. Error expressions during decoding In I2C decoding, when ACK (acknowledgement) is 1, ACK error occurs. When the decoding is correct, the ACK information is displayed as turquoise patch; when the detected ACK is 1, red error mark ( ;...
RIGOL Chapter 8 Protocol Decoding SPI Decoding (Option) The SPI serial bus is based on the master-slave configuration and usually consists of the chip select line (SS), clock line (SCLK) and data lines. Wherein, the data lines include the master input/slave output data line (MISO) and master output/slave input (MOSI) data line.
Page 253
Chapter 8 Protocol Decoding RIGOL setting. If the data channel in the trigger setting is an analog channel, the threshold level will also be copied. Note: After executing the copy operation, MISO Polarity in the decoding ― setting will be set to (positive polarity) automatically.
Page 254
RIGOL Chapter 8 Protocol Decoding Set the CS signal Press Channel and turn the multi-function knob to select D0 to D15 or CH1 to CH4. When an analog channel (CH1 to CH4) is selected, you can press Threshold and turn the multi-function knob to adjust the threshold (the default is 0 V) of the CS channel.
Page 255
Chapter 8 Protocol Decoding RIGOL no corresponding data channel is specified. 6. Specify the bits Press Data Bits and turn the multi-function knob to select the desired data bits. The range available is from 4 to 32 and the default is 8.
Page 256
RIGOL Chapter 8 Protocol Decoding When BusStatus is set to “ON”, press Event Table to enter the event table submenu. In this submenu, you can perform the following operations. Press Event Table to turn on or off the event table. When the event table is turned on, the interface is as shown in Figure 8-5.
Page 257
Chapter 8 Protocol Decoding RIGOL The decoded data information displayed in the bus is related to the horizontal time base. You can view the detailed information by reducing the horizontal time base. If a USB storage device is currently connected to the instrument, you can press Export to export the decoded data list to the external USB storage device in CSV format.
RIGOL Chapter 8 Protocol Decoding CAN Decoding (Option) The data frame format of the CAN bus is as shown in Figure 5-5. The ocillsocope samples the CAN signal on the specified sample position (if the source is an analog channel, the oscilloscope will also judge each data point (logic “1” or “logic “0”) according to the preset threshold level).
Page 259
Chapter 8 Protocol Decoding RIGOL 3. Specify the source channel Press Source; turn the multi-function knob to select the desired source channel and press the knob. You can also press Source continuously to switch the current source channel. You can select D0 to D15 or CH1 to CH4.
Page 260
RIGOL Chapter 8 Protocol Decoding adjust the signal rate at a relatively larger step. The adjustable range is from 10 kb/s to 1 Mb/s. Note: The larger the rotation amplitude of the outer knob of the navigation knob is, the faster the variation of the value will be. When you need to adjust the signal rate within a relatively larger range, this function can help you quickly finish the setting.
Page 261
Chapter 8 Protocol Decoding RIGOL to view relatively longer decoded data. Note: You can only view the event table and export the decoded data when the display of the corresponding bus is turned on. When BusStatus is set to “ON”, press Event Table to enter the event table submenu.
Page 262
RIGOL Chapter 8 Protocol Decoding Note: The display format of the decoded data in the event table is determined by the bus display format currently selected (refer to “Format”). When adjusting the horizontal time base, the waveform displayed on ...
Chapter 8 Protocol Decoding RIGOL FlexRay Decoding (Option) FlexRay is a differential serial bus configured with three continuous segments (packet header, payload and packet end). Its frame format is as shown in Figure 5-7. The oscilloscope samples the FlexRay signal on the specified sample position (if the source is an analog channel, the oscilloscope will also judge each data point (logic “1”...
Page 264
RIGOL Chapter 8 Protocol Decoding 3. Set the source channel Press Source; turn the multi-function knob to select the desired source channel and press the knob. You can also press Source continuously to switch the current source channel. You can select D0 to D15 or CH1 to CH4.
Page 265
Chapter 8 Protocol Decoding RIGOL 7. Adjust the sample position The sample point is a point within a bit’s time. The oscilloscope samples the bit levels at this point. The sample position is represented by the percentage of “the time from the start of the bit’s time to the sample point time” in the “bit’s time”, as shown in the figure below.
Page 266
RIGOL Chapter 8 Protocol Decoding Press Event Table to turn on or off the event table. When the event table is turned on, the interface is as shown in Figure 8-7. Number of Cycles Tail CRC Message ID Payload Length...
Page 267
Chapter 8 Protocol Decoding RIGOL horizontal time base. You can view the detailed information by reducing the horizontal time base. If a USB storage device is currently connected to the instrument, you can press Export to export the decoded data list to the external USB storage device in CSV format.
Chapter 9 Reference Waveform RIGOL Chapter 9 Reference Waveform MSO4000/DS4000 provides 10 reference waveform locations (namely Ref1 to Ref10). In actual testing process, users can compare the signal waveform with a reference waveform to judge the reason of the failure.
RIGOL Chapter 9 Reference Waveform To Turn the Reference Waveform Channel On/Off 1. Enable the reference waveform function Press REF in the vertical control area (VERTICAL) at the front panel to enable the REF function and open the REF menu. The backlight of the key turns on denoting that the reference waveform function is enabled.
Chapter 9 Reference Waveform RIGOL To Select the Current Reference Waveform Channel Press REF Current and turn the multi-function knob to select the desired reference waveform channel (Ref1 to Ref10). You can also press Current continuously to switch the current reference waveform channel.
RIGOL Chapter 9 Reference Waveform To Select the Reference Waveform Source Channel Press REF Current to select the specified reference waveform channel (Ref1 to Ref10) as the current reference waveform channel. Press Source and turn the multi-function knob to specify a source (CH1 to CH4, MATH or D0-D15) for the current reference waveform channel.
Chapter 9 Reference Waveform RIGOL To Adjust the Reference Waveform Display Display the reference waveform on the screen by referring to “To Save and Display the Reference Waveform”. Press and turn the multi-function knob to adjust the vertical position of the reference waveform.
RIGOL Chapter 9 Reference Waveform To Import from the Reference Waveform File Users can import the reference waveform stored in the internal memory or external USB storage device (only when a USB storage device is detected) to the current reference waveform channel.
Chapter 10 Pass/Fail Test RIGOL Chapter 10 Pass/Fail Test The Pass/Fail test function is used to monitor the change of the signal and judge whether the instrument passes or fails the test by judging whether the input signal is within the user-defined mask. This function is applicable to the analog channels (CH1 to CH4).
RIGOL Chapter 10 Pass/Fail Test To Enable or Disable the Pass/Fail Test Function Press Utility Pass/Fail to enter the test menu. In this menu, press Enable to enable or disable the pass/fail test function. Note: The pass/fail test function can not be enabled when the oscilloscope is in any of the following states.
Chapter 10 Pass/Fail Test RIGOL To Specify the Source Press Utility Pass/Fail Source and turn the multi-function knob to select the desired source channel. You can also press Source continuously to switch the current source channel. You can select CH1 to CH4.
RIGOL Chapter 10 Pass/Fail Test Screen Region Mode In the screen region mode, the test mask range is the whole careen region. Press Utility Pass/Fail Range to open the mask range menu. In this menu, 1. Specify the test mask range to the screen region Press Range to switch to “Screen”.
Chapter 10 Pass/Fail Test RIGOL Cursor Region Mode In the cursor region mode, you can set the test mask range by adjusting the positions of the cursors (cursor A and cursor B). Press Utility Pass/Fail Range to open the mask range menu. In this menu, 1.
Page 280
RIGOL Chapter 10 Pass/Fail Test 4. Adjust the vertical tolerance range Press Y Mask and turn the multi-function knob to adjust the vertical tolerance range of the test mask. The adjustable range is from 0.03 div to 4.00 div. During the adjustment, two white curves would be displayed to show the outline of the current mask.
Chapter 10 Pass/Fail Test RIGOL To Turn the Test Result Display On/Off Press Utility Pass/Fail Enable and select “ON” to enable the pass/fail test function; then, press Message to turn on or off the display of the test result statistic information.
Page 282
RIGOL Chapter 10 Pass/Fail Test [Trig Out/Calibration] connector at the rear panel each time a failed waveform is detected if AuxOutput is set to “ON”. 2. Turn the alarm on or off Press Output to select “Fail” or “Fail+ ”.
Chapter 10 Pass/Fail Test RIGOL To Save the Test Mask Range When the pass/fail test function is enabled, you can save the current test mask range to the internal memory or external USB storage device (only when a USB storage device is detected) in “*.pf”...
Chapter 11 Waveform Recording RIGOL Chapter 11 Waveform Recording The waveform recording function provides various recording modes (Record Mode). Record: record the channel waveforms at the specified interval until users stop the recording operation manually or when the number of frames recorded reaches the end frame currently set.
RIGOL Chapter 11 Waveform Recording To Record Waveform Using the “Record” Mode In the “Record” mode, the oscilloscope records the waveforms of all the channels currently turned on (include analog channels (CH1 to CH4) and digital channels (D0 to D15)) at the specified interval until users stop the recording operation manually or when the number of frames recorded reaches the end frame currently set.
Page 287
Chapter 11 Waveform Recording RIGOL Maximum Number of Frames In the waveform recording menu, Total Frames shows the the maximum number of frames that can be recorded currently. As the waveform memory capacity is fixed, more points per frame (namely the memory depth) results in a smaller number of waveform frames.
Page 288
RIGOL Chapter 11 Waveform Recording process. Press Interval and turn the multi-function knob or the inner knob of the navigation knob to adjust the interval at a relatively smaller step or turn the outer knob of the navigation knob to adjust the interval at a relatively larger step.
Page 289
Chapter 11 Waveform Recording RIGOL Stop automatically The oscilloscope will stop recording automatically after the specified number of frames (End Frame) are recorded. When the recording operation stops, the backlight of the stop shortcut key ) is illuminated in orange. When the recording operation is started again, the number of frames will start from 0.
RIGOL Chapter 11 Waveform Recording To Record Waveform Using the “Open” Mode In the “Open” mode, the oscilloscope records the waveforms of all the channels currently turned on (include the analog channels (CH1 to CH4) and digital channels (D0 to D15)) in real-time until users press RUN/STOP. The maximum number of frames can be recorded is limited by the memory depth and the on/off states of the channels.
Chapter 11 Waveform Recording RIGOL To Play Back the Waveform Recorded Waveform playback can play back the currently recorded waveform (using the “Record” or “Open” mode) in manual playback mode or auto playback mode. When the “Play back” mode is enabled, the real-time playback information as shown in the figure below is displayed at the upper-right side of the screen;...
RIGOL Chapter 11 Waveform Recording at the upper-right conrer of the screen changes accordingly. When the “Play back” mode is enabled, you can use SINGLE to play back the waveform manually. Each time SINGLE is pressed, the Current Frame moves one frame forward.
Page 293
Chapter 11 Waveform Recording RIGOL 3. Set the start frame Press Start Frame and turn the multi-function knob to set the start frame of playback. The maximum range is from 1 to the number of frames recorded. The default is 1.
Page 294
RIGOL Chapter 11 Waveform Recording 2) Pause the playback operation During the playback process (the backlight of the playback/pause shortcut ) is illuminated in yellow and flashes continuously), press Operate key ( and turn the multi-function knob to select “ ” and press the knob or directly press the playback/pause shortcut key ( ) at the front panel;...
Page 295
Chapter 11 Waveform Recording RIGOL When the playback operation stops, the backlight of the stop shortcut key ( is illuminated in orange. When the playback operation is started again, the current number of frame of the playback displayed at the upper-right side of the screen will start from the “Start Frame”.
RIGOL Chapter 11 Waveform Recording To Analyze the Recorded Waveform The waveform analysis function can analyze the analog channel waveforms recorded (use the “Record” mode or “Open” mode) and provide the analysis results. MSO4000/DS4000 supports trace and pass/fail test analysis modes.
Chapter 11 Waveform Recording RIGOL After waveform is recorded, press Record Mode; turn the multi-function knob to select “Analyze” and press the knob. At this point, the screen is divided into two display areas as shown in the figure below.
RIGOL Chapter 11 Waveform Recording Pass/Fail: perform analysis on the basis of the pass/fail test mask created by users. After selecting this mode, please set the desired test mask by referring to “To Create the Pass/Fail Test Mask (Pass/Fail Test Mode)”.
Page 299
Chapter 11 Waveform Recording RIGOL Note: The actual adjustable range of the start frame is from 1 to the end frame (End Frame) currently set. The larger the rotation amplitude of the outer knob of the navigation knob is, the faster the variation of the value will be.
Page 300
RIGOL Chapter 11 Waveform Recording During the waveform analysis, you can press Cancel to stop the analysis. Pressing Start again can restart the analysis. Note: During the analysis process, the progress bar is displayed on the screen and the settings cannot be modified.
Page 301
Chapter 11 Waveform Recording RIGOL under the pass/fail test analysis mode. 2) View the error frames or the specified frame of the current waveform analysis View the previous error frame Pressing Previous or RUN/STOP can locate the error frame previous to the “Current Error”...
Page 302
RIGOL Chapter 11 Waveform Recording 2. Set the specified waveform frame as the template Press Template to set the waveform frame currently displayed on the screen as the waveform analysis template. If the waveform frame currently displayed on the screen is not the one required,...
Page 303
Chapter 11 Waveform Recording RIGOL To Create the Pass/Fail Test Mask (Pass/Fail Test Mode) In the waveform recording menu (Mode is set to “Analyze”), press Analyze to select “Pass/Fail”. Then, set the test mask used in analysis based on Pass/Fail test mask through the method below.
Chapter 12 Display Control RIGOL Chapter 12 Display Control In the display menu (press Display at the front panel), you can set the type, persistence time and brightness of the waveform display as well as the grid type, grid brightness and the menu display time.
RIGOL Chapter 12 Display Control To Specify the Display Type Press Display to open the display menu and press Type to set the waveform display mode to “Vectors” or “Dots”. Vectors: the sample points are connected by lines and displayed. Normally, this ...
Chapter 12 Display Control RIGOL To Set the Persistence Time Press Display PersistTime and turn the multi-function knob to select the desired persistence time. You can also press PersistTime continuously to switch the current persistence time. You can select Min, specific values (from 50 ms to 20 s at 1-2-5 step) or Infinite.
Page 308
RIGOL Chapter 12 Display Control 3. Infinite In this mode, the oscilloscope displays the waveform newly acquired without clearing the waveforms acquired formerly. The waveforms acquired formerly will be displayed in relatively low-brightness color and the newly acquired waveforms will be displayed in normal brightness and color. Infinite persistence can be used to measure noise and jitter and to capture incidental events.
Chapter 12 Display Control RIGOL To Set the Waveform Brightness Press Display WaveIntensity and turn the multi-function knob to adjust the display brightness of the waveform . The range available is from 0% to 100%. The default is 50%.
RIGOL Chapter 12 Display Control To Set the Menu Display Time The menus at both sides of the screen will hold for a specified period of time (MenuDisplay) after the last button-pressing action and then are hidden. Press Display MenuDisplay and turn the multi-function knob to set the menu display time.
Chapter 13 Store and Recall RIGOL Chapter 13 Store and Recall Users can save the current settings, waveforms and screen image of the oscilloscope in internal memory or external USB mass storage device (such as a USB storage device) in various formats and recall the stored settings or waveforms when needed.
RIGOL Chapter 13 Store and Recall Storage System The internal memory (Local Disk) of MSO4000/DS4000 can store at most 10 setting files, 10 reference waveform files and 10 mask files of the Pass/Fail test. Besides, this oscilloscope provides two USB Host interfaces (one located on the front panel and another on the rear panel) used to connect USB storage devices for external storage.
Chapter 13 Store and Recall RIGOL Storage Type Press Storage Storage to select the desired storage type. The descriptions of all types are as follows. If no USB storage device is currently connected to the instrument, only “Setups” is available.
Page 314
RIGOL Chapter 13 Store and Recall 5. CSV You can save the data of the waveform displayed on the screen or the waveform data of the specified channels in the internal memory to the external memory in a single “*.csv” file. You can specify the filename and the saving directory and save the corresponding parameter file (*.txt) under the same directory using...
Chapter 13 Store and Recall RIGOL To Perform Internal Storage and Recall The internal memory supports the storage and recall of setup files, reference waveform files and pass/fail test mask files. In the following section, the internal storage and recall method and procedures are introduced by taking the setup file as an example.
Page 316
RIGOL Chapter 13 Store and Recall Note: Turn the multi-function knob to select and then press down the knob to return to the previous directory. In internal storage, New File and New Folder are not available. You can press REF Export to enter the disk management interface of reference waveform file storage.
Chapter 13 Store and Recall RIGOL To Perform External Storage and Recall Before performing external storage and recall, make sure that the USB storage device is connected correctly. The external memory supports the storage of all the types of files in Storage as well as the recall of “Traces”, “Waveform”...
Page 318
RIGOL Chapter 13 Store and Recall 4) Press Save to open the disk management interface. Turn the multi-function knob to select “Disk D” (the characters turn green) and press down the knob to open the external disk storage interface (as shown in Figure 13-2).
Page 319
Chapter 13 Store and Recall RIGOL same directory with the same filename as the picture file. Press Inverted to enable or disable the invert function. Press Color to set the color of the image to “GrayScale” or “Color”. Press Header to enable or disable the header display. When it is enabled, the header would display the instrument information, such as the model, date and time.
RIGOL Chapter 13 Store and Recall Disk Management Press Storage Disk.Manage to open the disk management interface as shown in the figure below. Turn the multi-function knob to select the desired disk (the disk currently selected is displayed in green) and press the knob to open the disk selected.
Chapter 13 Store and Recall RIGOL To Select File Type In addition to the file types in Storage, the oscilloscope can also display some files for advanced applications via the oscilloscope. Press Storage Disk.Manage File Type; turn the multi-function knob select the desired file type and press the knob.
RIGOL Chapter 13 Store and Recall To Create a New File or Folder This operation is only valid in external storage. Before performing the operation, make sure that a USB storage device is connected correctly. Press Storage Disk.Manage; turn the multi-function knob to select the external disk (“Disk D”...
Page 323
Chapter 13 Store and Recall RIGOL English Input Method The English input method interface is as shown in the figure below. Name Input Area Keyboard Upper-lower Case Switch Input Method Switch Figure 13-3 English Input Method Interface Name Input Area: display the characters currently input.
Page 324
RIGOL Chapter 13 Store and Recall and press the knob to switch to “ A”. Press Keyboard to select the “Keyboard” area, turn the multi-function knob to select “F” and press the knob to input the character. 4. Input the other characters Use the same method to input the remaining characters “ilename”...
Page 325
Chapter 13 Store and Recall RIGOL Chinese Input Method The Chinese input method interface is as shown in the figure below. Pinyin Input Area Chinese Character Selecting Area Input Method Switch Name Input Area Keyboard Upper-lower Case Switch Pinyin Input Area: display the Pinyin characters currently input.
Page 326
RIGOL Chapter 13 Store and Recall 3. Input the first Chinese character “文” Input the pinyin “wen” in the “Pinyin Input Area” Press Keyboard to select the “Keyboard” area, turn the multi-function knob to select the first letter “w” (or “W”) in the pinyin “wen” and press the knob to input “w”...
Chapter 13 Store and Recall RIGOL To Delete a File or Folder Folder operation is valid only in external storage mode. Before performing the operation, make sure that a USB storage device is connected correctly. 1. Delete a file in the internal memory 1) Press Storage ...
RIGOL Chapter 13 Store and Recall To Rename a File or Folder The rename operation is valid only in external storage mode. Before performing the operation, make sure that a USB storage device is connected correctly. Press Storage Disk.Manage; turn the multi-function knob to select the external disk (“Disk D”...
Chapter 13 Store and Recall RIGOL Factory Pressing Default OK can restore the oscilloscope to its factory state (refer to the table below). Table 13-2 Factory Parameter List Horizontal Setting (HORIZONTAL) Vertical Setting (VERTICAL) Acquisition Setting (Acquire) Trigger Setting (TRIGGER)
Page 330
RIGOL Chapter 13 Store and Recall CH3 Switch CH4 Switch Channel Coupling Bandwidth Limit Input Impedance 1 MΩ Channel Invert Vertical Scale Coarse Channel Unit Acquisition Setting (Acquire) Acquisition Mode Normal Sampling Mode Real Time Sa Rate 4.0 GSa/s La SaRate 1.0 GSa/s...
Page 331
Chapter 13 Store and Recall RIGOL Manual Display Type Source Select Select Cursor Time Unit Vertical Unit Source Unit Horizontal Positions of X Cursors CursorA -4*1 μs CursorB 4*1 μs Vertical Positions of Y Cursors CursorA 2*1 V CursorB -2*1 V...
Page 332
RIGOL Chapter 13 Store and Recall Network Configuration DHCP, Auto IP Mode USB Device Computer Sound Sound Pass/Fail Test Enable Test Source Operate Stop Mask Range Screen Region X Mask 0.24 div Y Mask 0.38 div Message Display Stop On Output...
Page 333
Chapter 13 Store and Recall RIGOL Vertical Expansion Ground Screen Saver Default Screen Saver Time External Trigger Impedance 1 MΩ Aux Output TrigOut Reference Clock ClockOutput Math Operation Setting (MATH Operation) Operate Source A Source B Invert Vertical Scale...
Page 334
RIGOL Chapter 13 Store and Recall Sample Rate 50 MSa/s Anti-Aliasing Logic Operation Expression Source A Source B Invert Vertical Scale Threshold A Threshold B Advanced Operation Expression Expression CH1+CH2 Invert Vertical Scale Logic Analyzer (LA) CH/Group On/Off D7-D0: Turn Off; D15-D8: Turn Off...
Page 335
Chapter 13 Store and Recall RIGOL Format Offset Threshold Parallel Clock Channel Slope Rising Edge Bus Bits Current Bit Channel RS232 None Polarity Normal Endian Baud 9600 bps Data Bits Stop Bit Even-Odd Check None Packet Packet End 00(NULL) SCLK...
Page 336
RIGOL Chapter 13 Store and Recall Source Signal Type Baud 500 kb/s Sample Point FlexRay Source Signal Path Signal Type Baud 10 Mb/s Sample Point Reference Waveform Setting (REF) Channel Setting REF1 Current Channel REF1 Source Color Gray Note : Only applicable to MSO4000 models.
Chapter 14 System Function Setting RIGOL Chapter 14 System Function Setting The contents of this chapter: Remote Interface Configuration System-related MSO4000/DS4000 User’s Guide 14-1 800.544.2843 www.calcert.com sales@calcert.co...
Chapter 14 System Function Setting Remote Interface Configuration MSO4000/DS4000 can communicate with PC via LAN, USB and GPIB (with the USB-GPIB interface converter provided by RIGOL) buses. Please refer to the introductions below to configure the corresponding interfaces before using the remote interfaces.
Page 339
Chapter 14 System Function Setting RIGOL Current Network Setting In the network connection setting interface, you can view the current network setting, including the network status, current IP configuration type, MAC address, VISA address and network parameters (IP address, subnet mask, default gateway and DNS server).
Page 340
RIGOL Chapter 14 System Function Setting MAC Address For each oscilloscope, the MAC address is unique. When attributing IP address for the oscilloscope, the MAC address is usually used to identify the instrument. VISA Address Display the current VISA address of the oscilloscope.
Page 341
Chapter 14 System Function Setting RIGOL To use Auto IP, turn on this configuration type and turn off the DHCP configuration type. To use Static IP, turn on this configuration mode and turn off the DHCP and Auto IP configuration types.
Page 342
RIGOL Chapter 14 System Function Setting The formats of the IP address, subnet mask, gateway address and DNS server address are all “nnn.nnn.nnn.nnn” which consists of 4 groups of nnn separated by “.”. Network Parameter Setting Method: Press the corresponding menu softkey to select the network parameter item to be set (by default, the first item in the address is selected).
Page 343
Chapter 14 System Function Setting RIGOL “Power-on Load” (Utility System Power On) is set to “Last” and “DHCP” and “Auto IP” are “Off” at the next power-on, the oscilloscope will load the preset IP address automatically. 5. Set the subnet mask (when Static IP configuration type is used) The format of the subnet mask is nnn.nnn.nnn.nnn;...
RIGOL Chapter 14 System Function Setting To Configure the USB Interface This oscilloscope can communicate with PC or PictBridge printer via the USB DEVICE interface at the rear panel. You need to set the oscilloscope to make it matchs with different device types.
Chapter 14 System Function Setting RIGOL System-related Sound When the beeper is enabled, you can hear the sound of the beeper when you operate a front panel key or menu softkey or when a prompt message pops up. Press Utility and press Sound to select (on) or (off).
RIGOL Chapter 14 System Function Setting System Time The system time is displayed at the lower-right side of the screen in “hh:mm (hour:minute)” format. When printing or storing a waveform, the file output will contain this time information. Press Utility System System Time to open the time setting menu. In this menu, 1.
Press Utility System Screen to open the screen saver setting menu. In this menu, Screensaver is “Default”; namely the icon RIGOL is used as the screen saver icon. Press Time and turn the multi-function knob to select the desired screen saver time (the default is “OFF”).
RIGOL Chapter 14 System Function Setting Error Information The failure that might occur includes the temperature of the mainboard is too high, the fan stops rotating, etc. If failure occurs, users need to return the instruments back to our factory for repair. At most 16 error messages lately appeared can be viewed.
Chapter 14 System Function Setting RIGOL Self-calibration The self-calibration program can make the oscilloscope quickly reach the optimum working state to acquire measurement values as accurate as possible. You can configure the instrument to perform the self-calibration program at any time; it is especially recommended when the temperature change is up to or more than 5℃.
RIGOL Chapter 14 System Function Setting Print Setting PictBridge is a new print standard. If your oscilloscope and printer both comply with the PictBridge standard, you can connect the oscilloscope to the printer using a USB cable to print the screen image directly. Devices complies with the PictBridge standard have the icon as shown in the figure on the right side.
Page 351
Chapter 14 System Function Setting RIGOL 4. Set the picture type Press File Type and turn the multi-function knob to select the desired picture type. You can also press this key continuously to switch the current picture type. The picture types include “Default”, “Exif/Jpeg” and “Bmp”.
RIGOL Chapter 14 System Function Setting 11. Resume the print When the print is paused, pressing Continue will resume the unfinished print. Note: This softkey is only available after the print is paused. 12. View the print status Press Status to open the PictBridge status information interface. You can view the current print statuses of the printer.
Chapter 14 System Function Setting RIGOL Aux Output Users can set the oscilloscope to output the specified type of signal from the [Trig Out/Calibration] connector at the rear panel. Press Utility AuxOutput and turn the multi-function knob to select the desired signal type.
RIGOL Chapter 14 System Function Setting Reference Clock This oscilloscope can output the internal 10 MHz sample clock signal from the [10MHz In/Out] connector at the rear panel as well as receive an external 10 MHz clock signal from the connector. This function can be used to synchronize multiple oscilloscopes.
Before installing an option, follow the steps below to generate the option license or option license file using the key according to your need. Log in RIGOL official website (www.rigol.com); click Customer Center License Generate to enter the software license generation interface.
Page 356
RIGOL Chapter 14 System Function Setting Press Editor to open the option license input interface as shown in the figure below. Option License Input Area Character Selecting Area Input the option license. Turn the multi-function knob to select the specified character and press the knob to input the character in the “Option License Input Area”.
Page 357
Chapter 14 System Function Setting RIGOL At this point, the prompt message “License file detected, install?” is displayed. Press License to start installing the option and the installation progress is displayed. After the option is installed, the current option interface is displayed.
Use PC software Users can use the PC software to send commands to remotely control the oscilloscope. Ultra Sigma of RIGOL is recommended. You can acquire Ultra Sigma from the resource CD in the standard accessories or download the latest version from RIGOL official website (www.rigol.com).
RIGOL Chapter 15 Remote Control Remote Control via USB 1. Connect the device Connect the oscilloscope (USB DEVICE interface) with your PC (USB HOST interface) using a USB cable. 2. Install the USB driver This oscilloscope is a USB-TMC device. After you connect the oscilloscope to the PC and turn on the oscilloscope for the first time (the oscilloscope is automatically configured to USB interface;...
Page 362
4. View the device resource The resources found will appear under the “RIGOL Online Resource” directory and the model number and USB interface information of the instrument will also 15-4 MSO4000/DS4000 User’s Guide 800.544.2843...
Page 363
Chapter 15 Remote Control RIGOL be displayed as shown in the figure below. 5. Control the instrument remotely Right click the resource name “MSO4054 (USB0::0x1AB1::0x04B1::DS4C0000000001::INSTR)” to select “SCPI Panel Control” to open the remote command control panel through which you can send commands and read data.
RIGOL Chapter 15 Remote Control Remote Control via LAN 1. Connect the device Connect the oscilloscope to your LAN using a network cable. 2. Configure the network parameters Configure the network parameters of the oscilloscope according to “To Configure the LAN Interface”.
Page 365
Note: If you want to remove the unwanted resource, click the resource name to select it and then click 4. View device resource The resources found will appear under the “RIGOL Online Resource” directory as shown in the figure below: 5. Control the instrument remotely Right click the resource name “MSO4054 (TCPIP::172.16.3.137::INSTR)”...
Page 366
RIGOL Chapter 15 Remote Control You can also load the LXI webpage by entering the IP address of the instrument in the address bar of the PC browser. 15-8 MSO4000/DS4000 User’s Guide 800.544.2843 www.calcert.com sales@calcert.co...
Chapter 15 Remote Control RIGOL Remote Control via GPIB 1. Connect the device Use the USB-GPIB interface converter to extend a GPIB interface for the oscilloscope. Then connect the oscilloscope to your PC using a GPIB cable. 2. Install the driver of GPIB card Install the driver of the GPIB card which has been connected to the PC correctly.
Page 368
normally. If not, please follow the corresponding prompt messages to solve the problem. 5. View device resource The GPIB resources found will appear under the “RIGOL Online Resource” directory, as shown in the figure below. 15-10 MSO4000/DS4000 User’s Guide 800.544.2843...
Page 369
Chapter 15 Remote Control RIGOL 6. Control the instrument remotely Right-click the resource name “MSO4054 (GPIB0::18::INSTR)” to select “SCPI Panel Control” to turn on the remote command control panel through which you can send commands and read data. MSO4000/DS4000 User’s Guide 15-11 800.544.2843...
Some commonly encountered failures and their solutions are listed below. When you encounter those problems, please solve them following the corresponding steps. If the problem remains still, please contact RIGOL and provide your device information (Utility System System Info).
Page 372
RIGOL Chapter 16 Troubleshooting can the waveform be displayed stably. (3) Check the trigger level: adjust the trigger level to the middle position of the signal. (4) Change the trigger coupling type: if edge trigger is currently selected, press MENU (in the trigger control area (TRIGGER) at the front panel) ...
Page 373
Chapter 16 Troubleshooting RIGOL connected to the oscilloscope and PC (or PictBridge printer). (3) If necessary, restart the oscilloscope. 9. The USB storage device cannot be recognized: (1) Check whether the USB storage device can work normally. (2) Make sure that the USB storage device being used is Flash type. This oscilloscope does not support hardware type USB storage device.
Chapter 17 Specifications RIGOL Chapter 17 Specifications All the specifications are guaranteed except the parameters marked with “Typical” and the oscilloscope needs to operate for more than 30 minutes under the specified operation temperature. Sample Sample Mode Real-time sample Real-time Sample...
Page 376
RIGOL Chapter 17 Specifications Input Number of MSO40X4: 4-analog-channel + 16-digital-channel Channels MSO40X2: dual-analog-channel + 16-digital-channel DS40X4: 4-channel DS40X2: dual-channel Input Coupling DC, AC or GND Input Impedance Analog channel: (1 MΩ±1%) || (15 pF±3 pF) or 50 Ω±1.5% Digital channel: (101 kΩ±1%) || (9 pF±1 pF) Probe Attenuation Analog channel: 0.01X to 1000X, in 1-2-5 step...
Page 377
Chapter 17 Specifications RIGOL Horizontal Time Base Scale MSO405X/DS405X: 1 ns/div to 1 ks/div MSO403X/DS403X: 2 ns/div to 1 ks/div MSO402X/DS402X: 2 ns/div to 1 ks/div MSO401X/DS401X: 5 ns/div to 1 ks/div Deviation between 1 ns (typical), 2 ns (maximum) Channels Max.
Page 378
RIGOL Chapter 17 Specifications Vertical Bandwidth (-3 dB) MSO405X/DS405X: DC to 500 MHz (50 Ω) MSO403X/DS403X: DC to 350 MHz MSO402X/DS402X: DC to 200 MHz MSO401X/DS401X: DC to 100 MHz Single Bandwidth MSO405X/DS405X: DC to 500 MHz (50 Ω) MSO403X/DS403X: DC to 350 MHz...
Page 379
Chapter 17 Specifications RIGOL 1 mV/div to 195 mV/div: ±0.1 div ± 2 mV ± 1.5% offset ESD Tolerance ±2 kV Channel to DC to maximum bandwidth: >40 dB Channel Isolation Vertical (Digital Channel) (MSO4000) Threshold adjustable threshold with 8 channels in 1 group Level Type TTL (1.4 V)
Page 380
RIGOL Chapter 17 Specifications Trigger CH1 to CH4: ±6 divs from center of the screen Trigger Level EXT: ±0.8 V Range Trigger Mode Auto, Normal, Single Holdoff Range 100 ns to 10 s High Frequency 50 kHz Rejection Low Frequency...
Page 382
RIGOL Chapter 17 Specifications Measure Cursor Manual Mode: voltage deviation between cursors (△V ), time deviation between cursors (△T), reciprocal of △T (Hz) (1/△T) Track Mode: voltage and time values of the waveform point Auto Mode: allow to display cursors during auto...
Page 385
Chapter 17 Specifications RIGOL General Specifications Probe Compensation Output Output Voltage About 3 V, peak-peak Frequency 1 kHz Power Power Voltage 100 to 127 V, 45 to 440Hz 100 to 240 V, 45 to 65Hz Power Maximum 120 W Fuse...
Chapter 18 Appendix RIGOL Chapter 18 Appendix Appendix A: Accessories and Options Description Order Number MSO4054 (500 MHz, 4+16 channels mixed MSO4054 signal oscilloscope) MSO4052 (500 MHz, 2+16 channels mixed MSO4052 signal oscilloscope) MSO4034 (350 MHz, 4+16 channels mixed MSO4034...
Page 388
Note : The bandwidth upgrade options can only be used to upgrade the specified low-bandwidth models to the corresponding high-bandwidth models. Note: All the options or accessories can be ordered from you local RIGOL office. 18-2 MSO4000/DS4000 User’s Guide 800.544.2843 www.calcert.com...
To get repair service, please contact with your nearest RIGOL sales or service office. RIGOL does not provide other warranty items except the one being provided by this warranty statement.
Index RIGOL Index - Duty ........6-25 End of Frame Error ....8-13 - Width ........6-25 Error Frames ......11-16 event table ........ 8-5 + Duty ........6-25 + Width ........6-25 Expression ........ 6-15 Acquisition Mode ......4-6 External Trigger Impedance ..
Page 392
RIGOL Index operation function ....6-14 subnet mask ......14-7 Operator ........6-17 sync trigger mode ..... 5-29 Overshoot ........ 6-29 Threshold ......... 6-30 Pass/Fail Test Mask ....10-3 time tag ......... 11-11 Pass/Fail Test Mode....11-12 TimeOut ........8-21 To Clear the Measurement ..
Page 393
Index RIGOL Waveforms ....... 13-3 X-Y Time Base Mode ....3-5 Window Function ......6-6 Y Cursor ........6-37 X Cursor ........6-37 Y-T Time Base Mode ....3-4 X-Y Cursor Mode ....... 6-53 zero offset ....... 2-17, 7-11 MSO4000/DS4000 User’s Guide 800.544.2843...