Measure:source; Measure:counter:source; Measure:counter:value - Rigol DS1000Z Series Programming Manual

Digital oscilloscope
Hide thumbs Also See for DS1000Z Series:
Table of Contents

Advertisement

RIGOL

:MEASure:SOURce

Syntax
:MEASure:SOURce <sour>
:MEASure:SOURce?
Description
Set or query the source of the current measurement parameter.
Parameter
Name
<sour>
Explanation
This command is used to set the source for the measurement parameters (except the
delay and phase). The delay and phase measurements requires two sources which can be
set by the
and
:MEASure:SETup:PSB
Return
The query returns CHAN1, CHAN2, CHAN3 or CHAN4.
Format
Example
:MEASure:SOURce CHANnel2
Related
:MEASure:SETup:PSA
Commands
:MEASure:SETup:PSB
:MEASure:SETup:DSA
:MEASure:SETup:DSB
:MEASure:ITEM

:MEASure:COUNter:SOURce

Syntax
:MEASure:COUNter:SOURce <sour>
:MEASure:COUNter:SOURce?
Description
Set or query the source of the frequency counter, or disable the frequency counter.
Parameter
Name
<sour>
Return
The query returns CHAN1, CHAN2, CHAN3, CHAN4 or OFF.
Format
Example
:MEASure:COUNter:SOURce CHANnel2
Related

:MEASure:COUNter:VALue?

Command
:MEASure:COUNter:VALue?
Syntax
:MEASure:COUNter:VALue?
Description
Query the measurement result of the frequency counter. The default unit is Hz.
Return
The query returns the measurement result in scientific notation. If the frequency counter
Format
is disabled, 0.0000000e+00 will be returned.
Example
:MEASure:COUNter:VALue?
Related
:MEASure:COUNter:SOURce
Command
2-54
Type
Range
Discrete
{CHANnel1|CHANnel2|CHANnel3|CHANnel4}
:MEASure:SETup:DSA
and :MEASure:SETup:DSB,
commands respectively.
Type
Range
Discrete
{CHANnel1|CHANnel2|CHANnel3|CHANnel4|OFF}
/*The query returns 1.000004e+03*/
Chapter 2 Command System
Default
CHANnel1
:MEASure:SETup:PSA
Default
OFF
DS1000Z Programming Guide

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ds1104zDs1074zDs1104z-sDs1074z-s

Table of Contents