Measure:statistic:reset; Measure:source - Rigol DS4000E Series Programming Manual

Digital oscilloscope
Hide thumbs Also See for DS4000E Series:
Table of Contents

Advertisement

RIGOL
Remarks
EXTRemum: The current value, average value, minimum value, and maximum value are displayed at
the bottom of the screen.
DIFFerence: The current value, average value, standard deviation, and count value are displayed at
the bottom of the screen.
Return Format
The query returns EXTR or DIFF.

:MEASure:STATistic:RESet

Syntax
:MEASure:STATistic:RESet
Description
Clears the history statistic data, makes statistics again, and displays the results.

:MEASure:SOURce

Syntax
:MEASure:SOURce <source>
:MEASure:SOURce?
Description
Sets or queries the source channel of the time, voltage, and area parameter measurement.
Parameter
Name
Type
<source>
Discrete
Remarks
When the math operation type is FFT (refer to the :CALCulate:MODE command), <source> cannot be
set to "MATH".
You need to specify two source channels for delay and phase parameter measurements. For the
source channel settings of delay parameter measurements, refer to the :MEASure:SETup:DSA
and :MEASure:SETup:DSB commands. For the source channel settings of phase parameter
measurements, refer to the :MEASure:SETup:PSA and :MEASure:SETup:PSB commands.
Return Format
The query returns CHAN1, CHAN2, CHAN3, CHAN4, or MATH.
2-154
Range
{CHANnel1|CHANnel2|CHANnel3|CHANnel4|MATH}
Chapter 2 Command System
Default
CHANnel1
DS4000E Programming Guide