7
Error Information
Detector Error Messages

Illegal Temperature Value from Sensor at Fan Assembly

Error ID: 1072
This temperature sensor (located close to the fan) delivered a value outside
the allowed range. The parameter of this event equals the measured
temperature in 1/100 centigrade. As a result the temperature control is
switched off.
Probable cause
1
2
3

Heater at fan assembly failed

Error ID: 1073
Every time the deuterium lamp or the tungsten lamp (DAD only) is switched
on or off a heater self-test is performed. If the test fails an error event is
created. As a result the temperature control is switched off.
Probable cause
1
2
130
The temperature sensor is defect.
Defective main board.
Detector is exposed to illegal ambient
conditions.
Defective connector or cable.
Defective heater.
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Verify that the ambient conditions are within
the allowed range.
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.