Generating Reports - Micromeritics Gemini VII 2390a Operator's Manual

Hide thumbs Also See for Gemini VII 2390a:
Table of Contents

Advertisement

Generating Reports

3.
Click Report after analysis to automatically generate reports when the analysis is complete. On
the Report Settings window, select the report destination. Click OK.
4.
Click Start to start the analysis. A window displays data as they are collected. A short delay is
encountered before the port status changes from the Idle state at the bottom of the screen
When analysis is complete, remove the sample tube and store (or dispose) of the sample material as
applicable.
Use caution when removing the sample tube if using a hanging filler rod. The sample
tube O-ring or Dewar cover may snag the filler rod retaining ring. Loosen the snag
gently; excessive force may break the tip of the filler rod.
Generating Reports
Reports > Start Report > [file] (or use the F8 keyboard shortcut)
1.
Select the report in the Name column. The selected report name appears in the File name text
box. To select more than one file, hold down the Ctrl key on the keyboard while selecting the
files. Click Open.
2.
Select the report destination in the Report Settings window and click OK. Refer to
Files, page 3-30.
3-28
Gemini VII
Printing
239-42828-01 - Dec 2012

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Gemini VII 2390a and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

This manual is also suitable for:

Gemini vii 2390tGemini vii 2390p

Table of Contents