Calibrating the System

Balance Servo

Use this option to calibrate the balance servo valve.
Ensure that empty tubes are installed on the sample and balance ports before
performing this procedure.
Ensure that empty tubes are installed on the sample and balance ports.
1.
Select Unit [n] > Calibration > Balance Servo.
2.
Click Start. The window closes when the calibration is complete.
9-32
Gemini VII
239-42828-01 - Dec 2012