Gemini VII
6049-
Operator cancelled operation.
Cause:
Action:
6050-
Valve failure.
Cause:
Action:
6051-
Power failure.
Cause:
Action:
6052-
Zero calibration failed.
6053-
Scale calibration failed.
6054-
Transducer match failed
6055-
Dosing servo calibration failed.
6056-
Balance servo calibration failed
6057
Quantity adsorbed calibration failed.
Cause A:
Action A:
Cause B:
Action B:
239-42828-01 - Nov 2013
Cancel was clicked while an analysis or a Po measurement was in
progress.
Restart the operation.
An error in the valve control electronics has occurred.
Contact a Micromeritics service representative if this error message
continues.
A power failure occurred during an automatic operation. The
operation was cancelled.
Restart the operation.
Calibration was out of the tolerance range.
Verify the preparatory steps for the calibration were performed
properly. For example, the sample tube may not have been empty.
If all preparatory steps were followed and no other message was
displayed, contact a Micromeritics service representative.
One of the steps in the calibration failed. An additional error message
is displayed.
Refer to the Cause/Action for the error message that displayed with
this one.
Appendix B
B-27