Gemini VII
Cause B:
Action B:
6039-
Po pressure underrange.
Cause:
Action:
6040-
Po pressure overrange.
Cause A:
Action A:
Cause B:
Action B:
6041-
Quantity adsorbed underrange.
Cause:
Action:
6042-
Quantity adsorbed overrange.
Cause A:
Action A:
Cause B:
Action B:
239-42828-01 - Nov 2013
An error occurred in the pressure measurement electronics.
Contact a Micromeritics service representative if this error message
continues.
A test of the Po pressure transducer showed a low reading. The
current operation will continue.
Contact a Micromeritics service representative if this error message
continues.
The gas pressure in the Po tube is too high. The current operation will
continue.
Verify that the gas inlet pressure regulator setting does not exceed 15
to 18 psi (103.4 to 124.1 kPa). Contact a Micromeritics service
representative if this error message continues.
An error occurred in the pressure measurement electronics.
Contact a Micromeritics service representative if this error message
continues.
A test of the quantity adsorbed pressure transducer showed a low
reading. The current operation will continue.
Contact a Micromeritics service representative if this error message
continues.
The rate of adsorption by the sample exceeded the response capacity
of the analyzer, preventing normal refilling of the gas reservoirs.
Repeat the analysis using a smaller amount of sample.
An error occurred in the pressure measurement electronics.
Contact a Micromeritics service representative if this error message
continues.
Appendix B
B-25