Connecting Gases
On the report, the Leak Rate field should read Passed. A Passed reading indicates the gas line
from the instrument to the regulator is clean and leak-free.
If the Leak Rate field indicates Failed, a leak is indicated between this connection. Check the
connections from the regulator shut-off valve to the gas bottle shut-off valve. Tighten as necessary,
then run the test again.
8.
Click the Gas Pressure Test - 3 tab to display its report.