Appendix B
6030-
A sample must be selected to proceed.
Cause:
Action:
6031-
Cannot read the analysis conditions parameter file
6032-
Cannot read the adsorptive properties parameter file.
6033-
Cannot read the report options parameter file.
Cause:
Action:
6035-
The instrument [S/N nn] is not calibrated.
Cause:
Action:
6036-
Pressure is out of range.
Cause:
Action:
6037-
Sample pressure underrange.
Cause:
Action:
6038-
Sample pressure overrange.
Cause A:
Action A:
B-24
Select a sample file prior to starting a blank analysis or a QuickStart
analysis.
Select a sample file and proceed with the analysis.
The file is corrupt.
Recreate the file or select a different file.
Upon initialization, the analyzer was unable to locate the calibration
file.
Select Unit > Calibration > Load from File and select a file
containing calibration data. Contact a Micromeritics service
representative if this error message continues.
There is a problem with the analyzer's calibration.
Select Unit > Calibration > Load from File and select a file
containing calibration data. Contact a Micromeritics service
representative if this error message continues.
A test of the sample pressure transducer showed a low reading. The
current operation will continue.
Contact a Micromeritics service representative if this error message
continues.
Adsorbate gas pressure is too high. The current operation will
continue.
Verify that the gas inlet pressure regulator setting does not exceed 15
to 18 psi (103.4 to 124.1 kPa). Contact a Micromeritics service
representative if this error message continues.
Gemini VII
239-42828-01 - Nov 2013