Micromeritics Gemini VII 2390a Operator's Manual page 289

Hide thumbs Also See for Gemini VII 2390a:
Table of Contents

Advertisement

Gemini VII
6025-
Serial write queue full.
Cause:
Action:
6026-
Measure Po for each point is not supported by this model. Choose another
sample file for analysis or change the Po option in the analysis conditions.
Cause:
Action:
6027-
Analysis: Po dosing failure
Cause:
Action:
6028-
This instrument does not provide measured Po analyses.
Cause:
Action A:
Action B:
Action C:
6029-
Instrument is not completely calibrated.
Cause:
Action:
239-42828-01 - Nov 2013
There was a fatal error in communication between the analyzer and
the software. All displays for that instrument will be closed.
Ensure the Ethernet cable is securely connected. Exit the application
and then restart it. Contact a Micromeritics service representative if
this error message continues.
Measure Po for each point was selected as the saturation method.
This option is only available for the Gemini 2390p or 2390t.
Edit the Po options in the analysis conditions and select Entered or
Most recent measured as the saturation pressure method.
.
The Dewar does not contain enough cryogen to take an accurate
reading in the Po tube.
Refill the Dewar with cryogen and restart the analysis.
An attempt was made to start an analysis on the Gemini 2390a using
a sample file in which Measure Po for each point was selected as the
saturation method. This feature is available only for the Gemini
2390p and 2390t.
Cancel the analysis, open the sample file, and choose Entered or
Most recent measured for the Po value. Restart the analysis.
Select a different file and continue with the analysis.
If a Gemini 2390p or 2390t is attached, select Start Analysis for that
unit.
The instrument in use is missing calibrations.
Select Unit > Calibration > Load from File and select a file
containing calibration data. Contact a Micromeritics service
representative if this error message continues.
Appendix B