  • Page 1 ® VII 2390 EMINI URFACE NALYZER NSTALLATION NSTRUCTIONS AND HECKLIST 239-42870-02 May 2021 Rev (B)
  • Page 2 Copyright The software described in this manual is furnished under a license agreement and may be used or copied only in accordance with the terms of the agreement. ____________________________________________________________________________ Copyright © 2009 - 2021. Micromeritics Instrument Corporation. All rights reserved.
  • Page 3: Warranty

    MICROMERITICS shall not be liable for consequential or other type damages resulting from the use of any of its products other than the liability stated above. This warranty is in lieu of all other warranties, express or implied, including but not limited to, the implied warranties of merchantability or fitness for use.
  • Page 4: Corporate Profile

    The company has R&D and manufacturing sites in the USA, UK, and Spain, and direct sales and service operations throughout the Americas, Europe, and Asia. Micromeritics systems are the instruments-of-choice in more than 10,000 laboratories of the world’s most innovative companies, prestigious government, and academic institutions. Our world-class scientists and responsive support teams enable customer success by applying Micromeritics technology to the most demanding applications.
  • Page 5: Contact Us

    Phone: 1-770-662-3636 Fax: 1-770-662-3696 www.Micromeritics.com Instrument Service or Repair Phone: 1-770-662-3666 International — contact your local distributor or call 1-770-662-3666 Service.Helpdesk@Micromeritics.com Micromeritics Learning Center Phone: 1-770-662-3607 www.Micro.edu Gemini VII 2390 Installation Instructions and Checklist 239-42870-02 Rev (B) — May 2021...
  • Page 6: About This Manual

    For instructions on installing the software in a 21CFR11 environment, see the Confirm Administrator Guide [part number 004-42821-01]. This document can be found on the Micromeritics web page (www.Micromeritics.com). All references to Gemini VII in this document encompass the Gemini VII models 2390a, 2390p, and 2390t unless otherwise noted.
  • Page 7: General Safety

    General Safety ENERAL AFETY Do not modify this instrument without the authorization of a Micromeritics service per- sonnel. Toxic or flammable gases require proper venting of exhaust. Any piece of laboratory equipment can become dangerous to personnel when improperly operated or poorly maintained. All employees operating and maintaining Micromeritics instruments should be familiar with its operation and should be thoroughly trained and instructed on safety.
  • Page 8 The instrument is intended to be used as per applicable local and national regulations. Gemini VII 2390 Installation Instructions and Checklist 239-42870-02 Rev (B) — May 2021...
  • Page 9 The Environmentally Friendly Use Period is valid only when the product is operated under the conditions defined in the product manual. Gemini VII 2390 Installation Instructions and Checklist viii 239-42870-02 Rev (B) — May 2021...
  • Page 10: Table Of Contents

    3 - 4 Reference Analysis 3 - 5 Keypad and Keyboard Functions 3 - 7 RS-232 Operation 3 - 9 Communications 3 - 10 Data Results 3 - 14 Gemini VII 2390 Installation Instructions and Checklist 239-42870-02 Rev (B) — May 2021...
  • Page 11 8 - 2 Software Installation in 21CFR11 Environments 8 - 5 Uninstall Software in 21CFR11 Environments 8 - 15 Maintaining Confirm Users and Groups 8 - 16 Gemini VII 2390 Installation Instructions and Checklist 239-42870-02 Rev (B) — May 2021...
  • Page 12 Users and Groups Worksheet for 21CFR11 10 - 2 Sample Data Worksheet for Gas Adsorption 10 - 3 Gemini VII 2390 EU Declaration of Conformity DoC - 1 Gemini VII 2390 Installation Instructions and Checklist 239-42870-02 Rev (B) — May 2021...
  • Page 13: Installation Overview

    1 Installation Overview NSTALLATION VERVIEW This document contains procedures to install and verify the operation of the Gemini VII 2390 analyzer (models 2390a, 2390p, and 2390t). Use these installation instructions and checklists to ensure that installation procedures are completed properly.
  • Page 14: Safe Servicing

    1 Installation Overview ERVICING Do not modify this instrument without the authorization of a Micromeritics service per- sonnel. To ensure safe servicing and continued safety of the instrument after servicing, service personnel should be aware of the following risks: Product specific risks that may affect service personnel: Electrical.
  • Page 15: Power

    Power OWER The Gemini VII 2390 is designed to operate with a universal input power supply (100-240VAC) at 50/60Hz. Noise-free power of the correct voltage and frequency, with a safety earth ground, should be available through a standard wall receptacle. The power outlet should be able to supply 15 amps @ 100 or 110VAC ±10% or 7.5 amps @ 230VAC ±10%.
  • Page 17: Equipment Installation

    The equipment must be positioned such that the mains supply is not obstructed and is easily accessible to disconnect the equipment from the AC main power supply. Gemini VII 2390 Installation Instructions and Checklist 2 - 1 239-42870-02 Rev (B) — May 2021...
  • Page 18: Vacuum Pump

    2 Equipment Installation ACUUM The Vacuum Pump Guide can be found on the Micromeritics web page (www.Micromeritics.com). Do not connect the equipment to the power source until the proper voltage selection has been made. Doing so could result in electrical shock and/or damage the assembly.
  • Page 19: Connect Ethernet Cable

    3. Use an Ethernet cable to connect the computer to the Ethernet switch. Do not use the uplink port. 4. Ensure the computer and the Ethernet switch are powered on. Do not use the uplink port when using an Ethernet switch. Gemini VII 2390 Installation Instructions and Checklist 2 - 3 239-42870-02 Rev (B) — May 2021...
  • Page 20: Connect The Power Cord

    Use of a power cord or power supply not provided with the instrument could cause personal injury or damage to the equipment. If a replacement is needed, contact your Micromeritics Service Representative. Detachable power supply cords with an inadequate rating could cause significant instrument damage or physical harm.
  • Page 21: Connect Optional Smart Vacprep To Analyzer

    The Smart VacPrep must be attached to an Ethernet switch. See Connect Ethernet Cable on page 2 - 3. The Smart VacPrep Operator Manual can be found on the Micromeritics web page (www.Micromeritics.com). Gemini VII 2390 Installation Instructions and Checklist 2 - 5...
  • Page 23: Keypad Only Installations

    Connect one end of the printer cable to one of the USB 2.00 connectors on the back of the analyzer. Connect the other end to the printer. Gemini VII 2390 Installation Instructions and Checklist 3 - 1 239-42870-02 Rev (B) — May 2021...
  • Page 24 4. Press CHOICE until Edit Setup is displayed. Then press ENTER. 5. Press CHOICE until Communications is displayed. Then press ENTER. 6. Obtain an IP address. Gemini VII 2390 Installation Instructions and Checklist 3 - 2 239-42870-02 Rev (B) — May 2021...
  • Page 25 3. Connect one end of the RS-232 cable to the connector labeled RS-232 on the back of the analyzer and the other end to the analytical balance. Gemini VII 2390 Installation Instructions and Checklist 3 - 3 239-42870-02 Rev (B) — May 2021...
  • Page 26: Operational Verification

    D. High pressure gauge E. Gas cylinder shut-off valve - CLOSED F. Regulator connecter nut G. Regulator control knob - OPEN H. Brass reducer fitting Gemini VII 2390 Installation Instructions and Checklist 3 - 4 239-42870-02 Rev (B) — May 2021...
  • Page 27: Reference Analysis

    Set Up group are appropriate for the reference material being analyzed. 10. At the prompts, verify or enter information at each prompt. 11. Press ENTER to start the analysis: Gemini VII 2390 Installation Instructions and Checklist 3 - 5 239-42870-02 Rev (B) — May 2021...
  • Page 28 If the analysis results are not within the tolerance level, repeat steps 2 through 10 to analyze the sample again. Verify that the value entered for the mass is correct. If the analysis results still fail to match, contact a Micromeritics Service representative. Gemini VII 2390 Installation Instructions and Checklist 3 - 6 239-42870-02 Rev (B) —...
  • Page 29: Keypad And Keyboard Functions

    Alt + 9 Ctrl + Q Begins a series of analyses QuickStart Alt + 5 Ctrl + R Reviews completed analysis data Review Gemini VII 2390 Installation Instructions and Checklist 3 - 7 239-42870-02 Rev (B) — May 2021...
  • Page 30 Ctrl + T Transmits analysis or calibration data over the Transmit serial line. Transmits a partial report if an auto- matic operation is in progress. Gemini VII 2390 Installation Instructions and Checklist 3 - 8 239-42870-02 Rev (B) — May 2021...
  • Page 31: Rs-232 Operation

    If the signals are correctly configured, contact the receiving device manufacturer for assistance. Gemini VII 2390 Installation Instructions and Checklist 3 - 9 239-42870-02 Rev (B) — May 2021...
  • Page 32: Communications

    DHCP address for the first time, this prompt displays: The system automatically tries to assign an IP address. If the system can determine an address, the address is displayed. Gemini VII 2390 Installation Instructions and Checklist 3 - 10 239-42870-02 Rev (B) — May 2021...
  • Page 33 This address is often the same as the instrument’s IP address with a 1 after the last dot instead of the last three characters [nnn.nnn.nnn.1]. Enter the Gateway address. Gemini VII 2390 Installation Instructions and Checklist 3 - 11 239-42870-02 Rev (B) — May 2021...
  • Page 34 Your IT Department may need to configure the server to accept e-mail from the analyzer. For proper configuration, the following information may be required: IP address for the Gemini Sender line: Micromeritics_2390 Gemini VII 2390 Installation Instructions and Checklist 3 - 12 239-42870-02 Rev (B) — May 2021...
  • Page 35 Stop Bits At the Parity prompt, select either None, Even, or Odd. Parity At the Xon/Xoff Protocol prompt, select either Disabled or Enabled. Xon/Xoff Protocol Gemini VII 2390 Installation Instructions and Checklist 3 - 13 239-42870-02 Rev (B) — May 2021...
  • Page 36: Data Results

    3. Press Enter on the keypad. The analyzer serial number displays in the window - for example, 107. 4. Click the serial number hyperlink. A last five analyses display in the window. Gemini VII 2390 Installation Instructions and Checklist 3 - 14 239-42870-02 Rev (B) — May 2021...
  • Page 37 5. Click the applicable file to view the data results. Or click the file for a current operation to monitor its progress. 6. Use the File > Print option to print a copy of the results. Gemini VII 2390 Installation Instructions and Checklist 3 - 15 239-42870-02 Rev (B) — May 2021...
  • Page 38: Adsorptive Line Test

    Informative prompt. Indicates how long a test will take or when an operator should open or close a valve. Read the message, then press ENTER. Gemini VII 2390 Installation Instructions and Checklist 3 - 16 239-42870-02 Rev (B) — May 2021...
  • Page 39 If Failed displays in the window, tighten all gas line connections for the adsorptive gas and restart the test. If Passed displays in the window, continue with the next step. Gemini VII 2390 Installation Instructions and Checklist 3 - 17 239-42870-02 Rev (B) — May 2021...
  • Page 40: Helium Line Test

    Read the message, then press ENTER. Operator action prompt. Instructs the operator to perform an action. Perform the action, then press ENTER. Gemini VII 2390 Installation Instructions and Checklist 3 - 18 239-42870-02 Rev (B) — May 2021...
  • Page 41 If Failed displays in the window, tighten all gas line connections for the adsorptive gas and restart the test. If Passed displays in the window, continue with the next step. Gemini VII 2390 Installation Instructions and Checklist 3 - 19 239-42870-02 Rev (B) — May 2021...
  • Page 43: Software Installation

    These prerequisites must be completed prior to installing the software: User Account Control in the Windows operating system must be enabled to ensure all com- ponents of the Micromeritics application are installed properly. If UAC is not enabled, right-click the setup.exe installer file and select Run as administrator.
  • Page 44 Install just for me. Only the person logged on to the computer at the time of installation will have access to the analyzer program. Gemini VII 2390 Installation Instructions and Checklist 4 - 2 239-42870-02 Rev (B) — May 2021...
  • Page 45 Select another folder from the list, then click Install. Enter a new folder name in the text box, then click Install. Gemini VII 2390 Installation Instructions and Checklist 4 - 3 239-42870-02 Rev (B) — May 2021...
  • Page 46 8. When installation completes, the Installed window displays. Click Next. 9. The Analyzer Configuration window is used to configure the system for each analyzer and serial number to be installed. Gemini VII 2390 Installation Instructions and Checklist 4 - 4 239-42870-02 Rev (B) — May 2021...
  • Page 47 Specify and enter the remaining portion of the the configured port is not listed IP Address, then click Next. The IP Address must be 192.168.77.100. Gemini VII 2390 Installation Instructions and Checklist 4 - 5 239-42870-02 Rev (B) — May 2021...
  • Page 48 [calibration] directory. Click Next. If not prompted for calibration files, skip this step. 12. On the Analyzer S/N [n] Installed window, click Next. A window displays for each unit installed. Gemini VII 2390 Installation Instructions and Checklist 4 - 6 239-42870-02 Rev (B) — May 2021...
  • Page 49 VacPreps on the Configure the Smart VacPreps window, then click Next. If not installing a Smart VacPrep, click Next. A Smart VacPrep can also be installed after the analyzer soft- ware installation is complete. Gemini VII 2390 Installation Instructions and Checklist 4 - 7 239-42870-02 Rev (B) — May 2021...
  • Page 50 I want to manually reboot later. 18. To create a desktop shortcut, select the Create a desktop icon checkbox. 19. Click Finish to close the setup program. Gemini VII 2390 Installation Instructions and Checklist 4 - 8 239-42870-02 Rev (B) — May 2021...
  • Page 51: Start The Application

    TART THE PPLICATION To start the application, either click the application shortcut on the desktop OR select the application from the Windows program list. Gemini VII 2390 Installation Instructions and Checklist 4 - 9 239-42870-02 Rev (B) — May 2021...
  • Page 53: Analyzer Preparation

    Use to specify how data should appear on the application windows and reports. This menu option is not available if using Restricted option presentation in a standard installation environment. Gemini VII 2390 Installation Instructions and Checklist 5 - 1 239-42870-02 Rev (B) — May 2021...
  • Page 55: Gas Connections

    Do not use flexible gas lines. Some flexible lines may appear to be appropriate, such as those with a herringbone covering, but the line may be coated internally with a polymer. Gemini VII 2390 Installation Instructions and Checklist 6 - 1...
  • Page 56 Ensure the gas cylinder is closed before connecting to the analyzer. Gemini VII 2390 Installation Instructions and Checklist 6 - 2 239-42870-02 Rev (B) — May 2021...
  • Page 57: Connect A Regulator And A Gas Line To A Gas Cylinder

    If pressure drops, repair the leak. 6. Connect the gas line to the regulator. a. Connect the gas line to the regulator connector. Gemini VII 2390 Installation Instructions and Checklist 6 - 3 239-42870-02 Rev (B) — May 2021...
  • Page 58 Open the gas cylinder shut-off valve and flow gas for 10 to 30 seconds. c. Close the gas cylinder shut-off valve and allow regulator pressure to go to zero. Gemini VII 2390 Installation Instructions and Checklist 6 - 4...
  • Page 59: Connect Gas To Analyzer

    2. Insert the gas line into the port and hand tighten the connector nut. Use a 7/16 in. (11 mm) wrench to tighten the nut until very snug. 3. If using helium, connect in the same manner. Gemini VII 2390 Installation Instructions and Checklist 6 - 5 239-42870-02 Rev (B) — May 2021...
  • Page 60: Clean And Verify The Gas Line

    1. Click the down arrow to the right of the Test field and select Gas Line Clean and Verify # Rev [n]. This procedure uses the analysis gas as an example. Select Clean and Verify Gemini VII 2390 Installation Instructions and Checklist 6 - 6...
  • Page 61 7. Click the Instrument to Gas Bottle Shut-off Valve - 2 tab to display its report. The Leak Rate field should read Passed indicating the gas line from the analyzer to the regulator is clean and leak-free. Gemini VII 2390 Installation Instructions and Checklist 6 - 7 239-42870-02 Rev (B) — May 2021...
  • Page 62 If running the test again, close the gas cylinder valve before starting the test. 9. Click Close to close the test report. Click Close again to close the test. Gemini VII 2390 Installation Instructions and Checklist 6 - 8 239-42870-02 Rev (B) — May 2021...
  • Page 63: Operational Verification

    7 Operational Verification PERATIONAL ERIFICATION Tools required for checking instrument calibration Tool Part Number 003-09622-10 Pressure Gauge 512-09800-00 Temperature Calibration Standard Gemini VII 2390 Installation Instructions and Checklist 7 - 1 239-42870-02 Rev (B) — May 2021...
  • Page 64: Measure Saturation Pressure

    Click Start. After analysis, go to Unit [n] > Unit Configuration. The recorded value now displays in the Last Measured Po field. Gemini VII 2390 Installation Instructions and Checklist 7 - 2 239-42870-02 Rev (B) — May 2021...
  • Page 65: Perform A Reference Material Analysis

    9. Fill the Dewar with liquid nitrogen to about 5 cm (2 in.) from the top. Allow the Dewar to equi- librate to ambient conditions (approximately 30 minutes). 10. Close the sample compartment door. 11. Go to Unit [n] > Start Analysis. Gemini VII 2390 Installation Instructions and Checklist 7 - 3 239-42870-02 Rev (B) — May 2021...
  • Page 66 The gas lines are not clean. Helium must be 99.995% pure. The measured free space is too high. This indicates the helium may not be pure enough. Gemini VII 2390 Installation Instructions and Checklist 7 - 4 239-42870-02 Rev (B) — May 2021...
  • Page 67: Clean And Label Sample Tubes

    4. Fill the sample tubes with warm water and place them in the ultrasonic cleaning unit, then place the filler rods in the unit. Turn on the ultrasonic cleaning unit for approximately 15 minutes. Gemini VII 2390 Installation Instructions and Checklist 7 - 5 239-42870-02 Rev (B) — May 2021...
  • Page 68 9. Remove the sample tubes and filler rods from the oven and allow to cool. Do not insert the filler rods at this time. Filler rods are inserted before the sample tube is installed on the analysis port. Gemini VII 2390 Installation Instructions and Checklist 7 - 6 239-42870-02 Rev (B) — May 2021...
  • Page 69 11. Rinse the sample tube closure with isopropyl alcohol, then wipe the sample tube closure dry with a clean, lint-free cloth. 12. Label the sample tube and stopper for identification. 13. Replace the rubber stopper. Gemini VII 2390 Installation Instructions and Checklist 7 - 7 239-42870-02 Rev (B) — May 2021...
  • Page 70: Create Sample Files

    The Degas Conditions tab displays only if enabled in Options > Option Presentation > Show Degas Conditions Gemini VII 2390 Installation Instructions and Checklist 7 - 8 239-42870-02 Rev (B) — May 2021...
  • Page 71 Description window. Use a mouse to click in the field first where information is to be entered then scan the bar code with the bar code reader. Gemini VII 2390 Installation Instructions and Checklist 7 - 9 239-42870-02 Rev (B) — May 2021...
  • Page 72: Determine The Sample Mass

    7. Remove the rubber stopper from the sample tube. 8. Use the sample tube funnel (provided in the accessories kit) and pour the sample from the weighing container into the sample tube. Gemini VII 2390 Installation Instructions and Checklist 7 - 10 239-42870-02 Rev (B) — May 2021...
  • Page 73: Degas Samples

    (www.Micromeritics.com). If using the Smart VacPrep degasser, use the Smart VacPrep tab in the application to set degas parameters. The Smart VacPrep Operator Manual can be found on the Micromeritics web page (www.Micromeritics.com). After degassing is complete, perform the following steps: 1.
  • Page 74: Sample Tube Installation

    It is not necessary to remove and replace the balance tube between analyses unless it has been contaminated or if using a different size sample tube. Gemini VII 2390 Installation Instructions and Checklist 7 - 12 239-42870-02 Rev (B) — May 2021...
  • Page 75 5. Place the Dewar cover under the sample and balance tubes and slide upward until it is 7.5 cm (3 in.) from connector nuts. Gemini VII 2390 Installation Instructions and Checklist 7 - 13 239-42870-02 Rev (B) — May 2021...
  • Page 76: Dewar Precautions

    If the Dewar has a protective mesh covering, do not remove it. This cover minimizes the risk of flying particles should the Dewar be knocked over, dropped, or broken. Gemini VII 2390 Installation Instructions and Checklist 7 - 14 239-42870-02 Rev (B) — May 2021...
  • Page 77: Fill And Install The Dewar

    4. Place the Dewar on the elevator. A Dewar support is not necessary for the larger Dewar used with the Gemini 2390t. Place the Dewar directly on the elevator. 5. Close the safety shield on the front of the analyzer. Gemini VII 2390 Installation Instructions and Checklist 7 - 15 239-42870-02 Rev (B) — May 2021...
  • Page 79: 21Cfr11 Environments

    If using a keypad with the analyzer, skip this chapter. VERVIEW The Micromeritics Confirm application enables laboratory managers in 21CFR11 environments to develop analysis methods, enforce industry standards, produce audit trails, and enables laboratory analysts to perform analyses and produce reports.
  • Page 80: User Permissions

    8 21CFR11 Environments ERMISSIONS The Confirm application uses Windows Users and Groups to control access to the Micromeritics Confirm application. When the Micromeritics Confirm application is installed, two Confirm groups are created. Confirm users must be assigned to only one group.
  • Page 81 Developer Group mic_[analyzer model number]_developer. Members of the Developer group: have rights to all functions of the Micromeritics application - includ- ing Advanced option presentation which allows the creation and modification of methods, sample files, and parameter files. can run an analysis.
  • Page 82 Can change system units and gas selections ü Browse outside of the Confirm installation location ü ü* Can add log entries * With administrative privileges only. Gemini VII 2390 Installation Instructions and Checklist 8 - 4 239-42870-02 Rev (B) — May 2021...
  • Page 83: Software Installation In 21Cfr11 Environments

    Do not proceed with installation until these tasks have been completed. b. If the prerequisites have been met, click Next. Gemini VII 2390 Installation Instructions and Checklist 8 - 5 239-42870-02 Rev (B) — May 2021...
  • Page 84 Install just for me. Only the person logged on to the computer at the time of installation will have access to the analyzer program. Gemini VII 2390 Installation Instructions and Checklist 8 - 6 239-42870-02 Rev (B) — May 2021...
  • Page 85 If the disk space is OK, click Next to accept the default installation directory, or click Browse change the installation location, then click Next. Gemini VII 2390 Installation Instructions and Checklist 8 - 7 239-42870-02 Rev (B) — May 2021...
  • Page 86 8 21CFR11 Environments 7. The Application User Configuration window is used to create the application controller and the two groups used by the Micromeritics Confirm application. Use the Users and Groups Worksheet for 21CFR11 on page 10 - 2 to determine group members. This worksheet may have been completed in the Pre-installation Instructions and Checklist document.
  • Page 87 Install, or enter a new folder name in the text box, then click Install Gemini VII 2390 Installation Instructions and Checklist 8 - 9 239-42870-02 Rev (B) — May 2021...
  • Page 88 In the Step 2 group box, enter the serial number for each analyzer to be attached to this computer. The serial number label is located either near the incoming power supply or near the ON/OFF switch. c. Click Next. Gemini VII 2390 Installation Instructions and Checklist 8 - 10 239-42870-02 Rev (B) — May 2021...
  • Page 89 Specify and enter the remaining portion of the the configured port is not listed IP Address, then click Next. The IP Address must be 192.168.77.100. Gemini VII 2390 Installation Instructions and Checklist 8 - 11 239-42870-02 Rev (B) — May 2021...
  • Page 90 15. On the Analyzer S/N [n] Installed window, click Next. A window displays for each unit installed. 16. The Product Release Information window contains information about updates and changes included in the release. Click Next. Gemini VII 2390 Installation Instructions and Checklist 8 - 12 239-42870-02 Rev (B) — May 2021...
  • Page 91 Click the appropriate link to register the product or to log in to the customer portal. After registration, click Next. b. To register later, click Next. Gemini VII 2390 Installation Instructions and Checklist 8 - 13 239-42870-02 Rev (B) — May 2021...
  • Page 92 19. In Windows Users and Groups, the Application Controller must be added to one of the Con- firm groups before the Application Controller can run the Confirm program. See Main- taining Confirm Users and Groups on page 8 - Gemini VII 2390 Installation Instructions and Checklist 8 - 14 239-42870-02 Rev (B) — May 2021...
  • Page 93: Uninstall Software In 21Cfr11 Environments

    To uninstall the software, double-click the uninstall.exe file located in the software installation directory, then follow the prompts. To uninstall the Micromeritics Confirm application, the owner of the application directory and its contents must be set to the account of the administrator that is removing the application installation.
  • Page 94: Maintaining Confirm Users And Groups

    Create User Accounts 1. Open the Users and Groups window. 2. In the first column, right-click the Users folder and select New User. Gemini VII 2390 Installation Instructions and Checklist 8 - 16 239-42870-02 Rev (B) — May 2021...
  • Page 95 Account is disabled 4. Click Create. Enter another user or click Close to close the window. Users may not display until the window is closed. Gemini VII 2390 Installation Instructions and Checklist 8 - 17 239-42870-02 Rev (B) — May 2021...
  • Page 96 User to a Group on the facing page. 4. Click Create. Enter another user or click Close to close the window. Users may not display until the window is closed. Gemini VII 2390 Installation Instructions and Checklist 8 - 18 239-42870-02 Rev (B) — May 2021...
  • Page 97 Maintaining Confirm Users and Groups Assign a User to a Group Upon installation of the Micromeritics Confirm application, the Analyst and Developer groups are created. Group names should not be changed, however, descriptions may be modified. See User Permissions on page 8 - 2 for a list of group permissions.
  • Page 98 7. Click Apply, then click OK. 8. Repeat this process until all applicable users have been added to each group. Gemini VII 2390 Installation Instructions and Checklist 8 - 20 239-42870-02 Rev (B) — May 2021...
  • Page 99: Installation Checklists

    Are the gas supply cylinders within 6 ft (2 m) of the analyzer inlet valves for standard installation? Line voltage has been verified to be within tolerances and changed if required. Gemini VII 2390 Installation Instructions and Checklist 9 - 1 239-42870-02 Rev (B) — May 2021...
  • Page 100 Ethernet cable from the computer to the Eth- ernet switch. Then connect the Ethernet based analyzers to the Ethernet switch. Install application software (not for keypad installations). Gemini VII 2390 Installation Instructions and Checklist 9 - 2 239-42870-02 Rev (B) — May 2021...
  • Page 101 (make / model / SN) Computer Operating System Printer (make / model / SN) Was the computer supplied by Micromeritics? Computer Installed By: Name: Date: Comments: Gemini VII 2390 Installation Instructions and Checklist 9 - 3 239-42870-02 Rev (B) — May 2021...
  • Page 102 Ethernet cable from the computer to the Ethernet switch. Then connect the Ethernet based analyzers to the Ethernet switch. Degasser Installation Evaluation Initial / Date Procedure Code Optional Smart VacPrep installed. Gemini VII 2390 Installation Instructions and Checklist 9 - 4 239-42870-02 Rev (B) — May 2021...
  • Page 103 Password configuration parameters entered. Expire: __________________________________ Warn: ___________________________________ Remember: ______________________________ Disable: _________________________________ Must match Windows user: __________________ Micromeritics application started. Logins tested for Administrator, Developer, and Analysis users. Gemini VII 2390 Installation Instructions and Checklist 9 - 5 239-42870-02 Rev (B) — May 2021...
  • Page 104 Gas lines attached to the instrument. Gas lines tested for leaks. Gas specified in application. Measure Saturation Pressure Evaluation Initial / Date Procedure Code Measure saturation pressure. Gemini VII 2390 Installation Instructions and Checklist 9 - 6 239-42870-02 Rev (B) — May 2021...
  • Page 105 The actual results must match the expected results, within the specified tolerance range, in order for this procedure to pass. Sample file name: (use the SN and your initials) __________________________________ Gemini VII 2390 Installation Instructions and Checklist 9 - 7 239-42870-02 Rev (B) — May 2021...
  • Page 106: Exceptions

    9 Installation Checklists XCEPTIONS Explain the procedures that were categorized as Fail or No. Comment Initial / Date Procedure Gemini VII 2390 Installation Instructions and Checklist 9 - 8 239-42870-02 Rev (B) — May 2021...
  • Page 107: Signatures

    Service Support Center in Norcross, GA, USA. After completing the installation process, representatives of Micromeritics who install analyzers are required to send the following documents to Micromeritics for inclusion in the customer's analyzer history:...
  • Page 109: Worksheets

    10 Worksheets 10 W ORKSHEETS Sample Data Worksheet for Gas Adsorption on page 10 - 3 Users and Groups Worksheet for 21CFR11 on the next page Gemini VII 2390 Installation Instructions and Checklist 10 - 1 239-42870-02 Rev (B) — May 2021...
  • Page 110: Users And Groups Worksheet For 21Cfr11

    _______________________________________________ User Name □ □ _______________________________________________ Full Name _______________________________________________ User Name □ □ _______________________________________________ Full Name _______________________________________________ User Name □ □ _______________________________________________ Full Name Gemini VII 2390 Installation Instructions and Checklist 10 - 2 239-42870-02 Rev (B) — May 2021...
  • Page 111: Sample Data Worksheet For Gas Adsorption

    Mass of sample (step 2 minus step 1) Degas Information Degas apparatus Temperature (°C) Time (hours) Actual time started Actual time finished Degas Notes: Gemini VII 2390 Installation Instructions and Checklist 10 - 3 239-42870-02 Rev (B) — May 2021...

This manual is also suitable for: