Micromeritics Gemini VII 2390a Operator's Manual page 418

Hide thumbs Also See for Gemini VII 2390a:
Table of Contents

Advertisement

6 Maintenance and Troubleshooting
3. Remove and discard the exhaust filter. Ensure the centering ring remains in place on the exhaust
port.
4. Place the new filter on the centering ring.
5. Open the clamp and place it around the exhaust port flange and the exhaust filter flange. Swing
the clamp fastening screw toward the exhaust port until it fits into the slot in the other half of the
clamp. Tighten the wing nut securely by hand.
6. Turn on the power to the pump.
Related Topics
See "Inspect and Change Vacuum Pump Oil" on page 6 - 3
See "Replace the Alumina in the Oil Vapor Trap" on page 6 - 7
6 - 12