Set File Type And Device - Keysight N9923A FieldFox User Manual

Hide thumbs Also See for N9923A FieldFox:
Table of Contents

Advertisement

State Files
ALL FieldFox instrument settings for ALL Modes are saved and recalled in a *.sta file.
If error correction is ON in CAT or NA Mode, then calibration data is also saved and
recalled.
If error correction is OFF (CalRdy is active), calibration data is NOT saved in the *.sta
file. When a Trace + State file is recalled on a FieldFox other than the one it was
saved on, differences in the CalRdy calibration data may result in differences
between the displayed traces.
Image Files
The FieldFox can open and display *.png, *.jpg, and *.jpeg files.
How to Recall an *.sta File
NOTE
Before recalling a file, first set File Type to State or Trace + State, and also Select
Device from which the file will be recalled.
 Press Save/Recall 9.
 Then Recall.
 Navigate to the file to be recalled using the Up/Down arrow keys.
 Then
Set File Type and Select Device
The following File TYPE and DEVICE settings are used to determine the type of file to
save, and the location to be used for all file operations.
These two settings remain through an instrument preset, but are reset to the default
values when the FieldFox power is shutdown.
Select Device (location of File Save/Recall)
 Press Save/Recall 9.
 Then
NOTE
Insert only ONE USB Flash Drive in the FieldFox when saving or copying files.
How to set File Type
State files are the only file type that can be recalled by the FieldFox.
 Press Save/Recall 9.
 Then
126
or Enter.
Recall File
to set the LOCATION where the file is to be saved. Choose
Select Device
from the following:
Saves/Recalls files to/from the FieldFox internal memory. To specify a
Internal
o
folder, first select Manage Folders (learn how on page 128). The Data Link
software will read data ONLY from the \UserData folder.
Saves/Recalls files to/from the micro SD card (not included), inserted
SD Card
o
in the card slot. See FieldFox Tour on page 15. Only the root folder is available.
Saves/Recalls files to or from a USB Flash drive (not included) that is
USB
o
inserted in one of the USB connectors. Only the root folder is available.
Type. Choose one of the following:
File
Saves ALL FieldFox instrument settings for ALL Modes to a *.sta file. If a
State
o
calibration is ON in CAT or NA Mode, calibration data is also saved. This file can
be opened ONLY by a FieldFox.
FieldFox User's Guide

Advertisement

Table of Contents
loading

Table of Contents