Data Sheet For Enclosure/Chassis Leakage Current - GE VIVID S60N Basic Service Manual

Hide thumbs Also See for VIVID S60N:
Table of Contents

Advertisement

P R E L I M I N A R Y
D
FR091521, R
IRECTION
NOTE:
For more information, refer to the safety analyzer's user manual that will be used to perform the tests."
The maximum allowable limit for chassis source leakage is shown in:
10-6-6-3

Data Sheet for Enclosure/Chassis Leakage Current

Table 10-15
Measurements should be recorded from multiple locations for each set of test conditions.
The actual location of the test probe may vary by Ultrasound system.
NOTE:
Values in italics font are given as examples only.
Record all data in the Electrical safety tests log.
Table 10-15 Typical Data Format for Recording Enclosure/Chassis Leakage
Unit under test____________________________________
Test Conditions
System
Power
off
off
on
on
NOTE:
Values in italics font are given as examples only.
1
EVISION
Table 10-11
on page 10-17
or
Table 10-12
on page 10-17
as Chassis/Enclosure Leakage.
below shows a typical format for recording the enclosure/chassis leakage current.
Grounding/
PE
closed
open
closed
open
Chapter 10 - Care and Maintenance
V
IVID
Measurement/Test Point Location
Rear
Lower
Panel
Frame
S60N/V
S70N
B
IVID
ASIC
Date of test:_____________
Probe
Connector
Main Handle
S
M
ERVICE
ANUAL
10-21