Data Sheet For Chassis Source Leakage Current - GE Voluson 730Pro Service Manual

Hide thumbs Also See for Voluson 730Pro:
Table of Contents

Advertisement

GE H
- K
EALTHCARE
RETZTECHNIK
D
, R
IRECTION
EVISION
10-7-5-3

Data Sheet for Chassis Source Leakage Current

The test passes when all readings measure less than the value shown in
Record all data on the Ultrasound Inspection Certificate.
Table 10-14 Typical Data Sheet for Chassis Source Leakage Current
Unit Power
ON
ON
ON
ON
OFF
OFF
OFF
OFF
10-7-6
Isolated Patient Lead (Source) Leakage–Lead to Ground
10-7-6-1
Definition
This test measures the current which would flow to ground from any of the isolated ECG leads. The
meter simulates a patient who is connected to the monitoring equipment and is grounded by touching
some other grounded surface. Measurements should be made with the ground open and closed, with
power line polarity normal and reversed, and with the ultrasound console Off and On. For each
combination the operating controls, such as the lead switch, should be operated to find the worst case
condition.
Equipment damage possibility. Never switch the Polarity when the unit is powered ON.
CAUTION
Be sure to turn the unit power OFF before switching the polarity using the POLARITY switch.
Otherwise, the unit may be damaged.
!! CAUTION:
DRAFT (A
Tester Neutral
Tester Polarity
Switch
Enter Name of tested peripheral here:
NORM
NORM
REV
REV
NORM
NORM
REV
REV
Chapter 10 - Care & Maintenance
28, 2008)
PRIL
Test 1
or Ground
Probe
Switch
Connector
OPEN
CLOSED
OPEN
CLOSED
OPEN
CLOSED
OPEN
CLOSED
V
® 730P
OLUSON
Table
10-11.
Test 2
Test 3
Optional
Wheel
CRT
Test 4
/ P
V (BT03)
RO
RO
S
M
ERVICE
ANUAL
Optional
Test 5
10-17