Dismount of the valves group
Dismount the eventual circuits or bags
connected to the valves group
connectors.
Pull up the valves group cover.
Pull up the handle.
Extract the valves group, pulling it up
vertically.
The pneumatic connections are
automatic with double tightness
gaskets to prevent accidental leakages.
WARNING. Risk of injury for the user/patient
All the interventions must be effected exclusively by personnel highly
qualified and specifically trained and formally authorized by SIARE.
WARNING. Risk of malfunctioning
To be effected only in case of maintenance on valves group. Pay much
ATTENTION during this operation.
Lift and remove completely the valves
group and position it on a flat surface.
Continue with the maintenance (cfr.
10).
User's Manual, DU3310103-MRI / rev.3 - 02.05.2019
Valves group
3-9