Ventilation Parameters Setting Area - Siare Morpheus Series User Manual

Anesthesia unit
Hide thumbs Also See for Morpheus Series:
Table of Contents

Advertisement

AD.A module

5.2.5 Ventilation parameters setting area

Trigger
Flow
Vt
Pmax
PEEP
5-14
Set the flow level (pressure) to recognize the patient spontaneous
activity. (l/min) (cmH 2 O).
From OFF, 1 to 15 cmH 2 O / From OFF, -1 to -9 l/min
Upon selected operative mode, with this parameter it is possible to
set the inspiratory flow value to be delivered by the ventilator. The
inspiratory flow value can be set in PCV modality only.
From 1 to 80 l/min.
Upon selected operative mode, with this parameter it is possible to
set the value of tidal volume delivered by the ventilator.
From 50 to 1500 ml (set value in function of a max Flow value of 80
liters).
Set the max limit value of airways pressure.
From 10 to 80 cmH2O
Set the value of airways positive pressure in expiratory phase.
From OFF, 3 to 30 cmH 2 O
MORPHEUS MRI