Alarms: AD.A module
FiO2
Apnoea
Vte
8-6
Low oxygen
concentration
Operating or not present
cell
Apnoea
Expired tidal volume
The flow sensor calibration is to be performed when there are
differences higher than 20% between the set volume (Vt) and
measured value of expired tidal Volume (Vte).
Fixed value: it is activated when the
oxygen concentration is lower and/or
equal to 19%
It is activated when the cell is exhausted
or disconnected: no signal is detected by
the system.
It is activated when the RATE value is
equal to 0; when there is no breathing
activity of the patient.
It is activated when the value of expired
tidal value is higher (lower) than value in
% of maximum (minimum) limit set.
Set default value ± 50%.