Alarms: AD.A module
8.5
Description
Gas Supply
High PAW
Low PAW
FiO2%
Apnoea
8-10
It is activated if the oxygen supply pressure is at a value lower
than minimum required for equipment operation.
It is signalled by an acoustic and visual alarm.
This alarm indicates the violation of upper limit (set in the
operative mode) of airways pressure.
It is activated once the alarm level of inspiratory pressure is
reached: in such case the expiratory valve will be opened
immediately.
It is signalled by an acoustic and visual alarm.
This alarm indicates the violation of lower limit (preset in the
automatic or manual operative mode at 5 cmH 2 O) of airways
pressure. It is an alarm that is activated in the automatic and
manual modalities after 20 s.
The low pressure alarm limit is meant respect to set PEEP
value.
It is signalled by an acoustic and visual alarm. It can be
silenced (suspended) for 20 s.
This alarm signals the overcome of lower limit of
corresponding parameter (19%).
It is signalled by an acoustic and visual alarm. It can be
silenced (suspended) for 20 s.
This alarm is activated when the value of RATE is equal to 0
and there is no patient breathing activity.
It is signalled by an audible and visual alarm. It can be
silenced (suspended) for 20 s.
MORPHEUS MRI