1. Press for around 5 seconds the soft key of ventilation parameters monitoring.
2. The flow sensor calibration procedure starts:
The calibration phases are 5, and they corresponds to 5 flow values delivered
o
by MORPHEUS: 0.0, 2.5, 5.0, 10.0, 20.0 l/min.
During the calibration phase the valves group bellow rises and falls
o
depending on the delivered flow.
3. In the parameters monitoring area is displayed the acronym "
indicates that the flow sensor calibration phase is active.
User's Manual, DU3310103-MRI / rev.3 - 02.05.2019