Preparation to use
BY-PASS push button.
•
Close all three regulators of the flowmeter
box.
•
Enabled selection control for valves group
enabling (9.4).
•
Press the BY-PASS (9.2) push button.
•
Verify that the O 2 flow reaches the valves
group bag.
•
Enabled fresh gas exit selection control (9.3)
•
Press the BY-PASS (9.2) push button
•
Verify that the O 2 flow reaches the TO and
FRO circuit bag connected to the AUX
connector (9.5).
•
Restore the control enabling the selection of
valves group (9.4).
•
Apply the patient circuit to valves group;
INSP.and EXP. connectors
•
Apply the patient simulator (SIARE type) to
the patient circuit connector.
6-42
This function is active also with anaesthesia unit OFF.
The suggested patient simulator for tests and checks is the
LS.AB.001 which is equipped with variable compliance and
resistance.
MORPHEUS MRI