Low battery
(10 minutes)
Disconnected
/ exhausted
O
cell
2
FiO 2 %
User's Manual, DU3310103-MRI / rev.3 - 02.05.2019
This alarm condition is present when the charge battery level is such
to be guaranteed a residual autonomy of about 10 minutes.
•
Verify the correct connection of power supply.
•
To recharge the battery.
•
If the alarm is activated when the battery has not provided the time
autonomy indicated on the technical sheet, request the intervention of a
Service Centre.
This alarm condition indicates the connection status of the oxygen
sensor or when the reagents are exhausted.
•
Check that the oxygen cell is correctly connected.
•
Replace the oxygen sensor with a new one.
•
Check the condition of the cable and the connector (if necessary, restore
the connection and replace the cable if damaged).
•
If the problem persists, contact the Siare Service Centre or a Centre
authorised by Siare.
This alarm is activated when the measured concentration of oxygen is
below the set limit.
•
Check that the oxygen cell is fitted correctly in its housing.
•
Calibrate the oxygen cell: if the problem occurs again after a short time,
replace the oxygen cell.
•
Check that the feeding pressure of the medical gases is correct: if it is
not, check the pressure of the distribution system and the correct
connection to the supply.
•
Check that the mask, endotracheal tube and patient circuit are not in
some way clogged, bent or crushed. If this is the case, eliminate the
problem or replace them.
•
Contact the Siare Service Centre or a Centre authorised by Siare.
Troubleshooting
9-3