Siare Morpheus Series User Manual page 182

Anesthesia unit
Hide thumbs Also See for Morpheus Series:
Table of Contents

Advertisement

Low battery
(10 minutes)
Disconnected
/ exhausted
O
cell
2
FiO 2 %
User's Manual, DU3310103-MRI / rev.3 - 02.05.2019
This alarm condition is present when the charge battery level is such
to be guaranteed a residual autonomy of about 10 minutes.
Verify the correct connection of power supply.
To recharge the battery.
If the alarm is activated when the battery has not provided the time
autonomy indicated on the technical sheet, request the intervention of a
Service Centre.
This alarm condition indicates the connection status of the oxygen
sensor or when the reagents are exhausted.
Check that the oxygen cell is correctly connected.
Replace the oxygen sensor with a new one.
Check the condition of the cable and the connector (if necessary, restore
the connection and replace the cable if damaged).
If the problem persists, contact the Siare Service Centre or a Centre
authorised by Siare.
This alarm is activated when the measured concentration of oxygen is
below the set limit.
Check that the oxygen cell is fitted correctly in its housing.
Calibrate the oxygen cell: if the problem occurs again after a short time,
replace the oxygen cell.
Check that the feeding pressure of the medical gases is correct: if it is
not, check the pressure of the distribution system and the correct
connection to the supply.
Check that the mask, endotracheal tube and patient circuit are not in
some way clogged, bent or crushed. If this is the case, eliminate the
problem or replace them.
Contact the Siare Service Centre or a Centre authorised by Siare.
Troubleshooting
9-3