At the end of calibration procedure, if all has been performed properly, a
FiO
For a correct calibration it is necessary that the cell is mounted in the
apposite location and electrically connected by the apposite cable.
If one of these conditions have not been effected the calibration is not
correct.
REPLACEMENT OF OXYGEN CELL
The oxygen cell must be replaced when at the end of calibration phase
the measured value is indicatively lower than 20% and/or in case of
displayed alarm message.
For ordering the spare cell and for disposal of depleted one consult the
"Maintenance" chapter.
User's Manual, DU3310103-MRI / rev.3 - 02.05.2019
percentage value equal to 99% should appear .
2
At the end of calibration procedure a FiO
value equal to 99% is displayed.