WARNING! Collision danger
Pay much WARNING during moving, positioning and use phases.
The devices or accessories applied can collide with other objects or
persons present in the path and/or in the operating room.
WARNING! Accidental moving danger
If the anaesthesia unit is not correctly positioned, it could accidentally
move during operation.
WARNING! Risk of injury for the user / patient
User's Manual, DU3310103-MRI / rev.3 - 02.05.2019
Position correctly the anaesthesia unit on a flat surface.
Apply the anaesthesia unit brakes to ensure that it could not
accidentally move during operation.
The assembly and connection of all the accessories must be
carried out by highly qualified technical personnel, trained and
formally authorised by SIARE.
This type of anaesthesia unit is not suitable for and therefore
cannot be used in a hyperbaric chamber.
To avoid an increase in the concentration of oxygen in the
surrounding air, the anaesthesia unit should only be used in
appropriately ventilated rooms.
Do not connect or disconnect parts or components when the
anaesthesia unit is on or connected to the mains power supply.
Before using the anaesthesia unit, carry out all the necessary
preliminary tests.
Preparation to use
6-3