AD.A module
Trigger
The threshold of flow level is set (or pressure) for recognition of patient
spontaneous activity in l/min (cmH 2 O).
The ventilator. once recognized such spontaneous activity provides to
deliver a synchronized inspiratory act with patient tented breath
The detection of spontaneous activity is displayed when the orange
indicator lights on.
Flow
Upon selected operative mode, with this parameter it is possible to set
the inspiratory flow value to be delivered by the ventilator.
The value of inspiratory flow can be set in PCV modality only.
Vt
Upon selected operative mode, with this parameter it is possible to set
the value of tidal volume delivered by the ventilator.
The tidal volume value, can be set in VAC modality only.
7-22
Pressure Trigger from: OFF, 1 to 15 cmH 2 O.
Flow Trigger from: OFF, 1 to -9 l/min.
Set flow from: 1 to 80 l/min.
Set tidal volume from: 50 to 1500 ml (value which can be set in
function of max flow value of 80 liters).
MORPHEUS MRI