Flowmeter Box - Siare Morpheus Series User Manual

Anesthesia unit
Hide thumbs Also See for Morpheus Series:
Table of Contents

Advertisement

7.2

Flowmeter box

7.2.1 Dosing and administration of fresh gas
The term " fresh gas " means the gas mixture that is dosed in the anaesthesia
module and in the anaesthetic vaporizer
The fresh gas is continuously delivered to the valves group to be administered to the
patient.
The excess of gas that may be present in the patient circuit is automatically
eliminated through the scavenger connection.
The quantity (in l/min) of fresh gas can be greater, equal to or lower than the Minute
Volume:
if it is higher or equal it is an high flow open circuit (without soda lime
canister)
if it is lower it is a semi-closed circuit
if the flow is lower than 1 l/min it is a low flow semi-closed circuit
WARNING! Risk of injury to the user / patient
Administration of fresh gas in the breathing system
In order to deliver the fresh gas flow into the valves group (breathing
system) press the APL key on the work shelf.
Otherwise, pressing the AUX key the fresh gas will go onto the TO and
FRO.
User's Manual, DU3310103-MRI / rev.3 - 02.05.2019
AD.A module
7-3